F4294-09 Hamamatsu Corporation, F4294-09 Datasheet

no-image

F4294-09

Manufacturer Part Number
F4294-09
Description
MCP ASSEMBLY WITH CENTER HOLE FOR REFLECTRON MS
Manufacturer
Hamamatsu Corporation
Datasheet
The F4294-09 is an MCP assembly developed for reflectron mass
spectrometers. In spite of having a center hole for beam line, the
F4294-09 eliminates ringing phenomena which have been a
problem with conventional MCP assemblies with a center hole.
The F4294-09 thus makes high-speed ion detection possible.
This allows a significant size reduction in the TOF mass spec-
trometer and a dramatic improvement in mass resolution.
FEATURES
SPECIFICATIONS
MAXIMUM RATINGS (Absolute values)
* Baking cannot be performed.
CHARACTERISTICS (at 1.3 10
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject
to change without notice. No patent rights are granted to any of the circuits described herein. © 1998 Hamamatsu Photonics K.K.
GENERAL
Assembly Outer Diameter
Effective Diameter
Maximum Height
MCP Channel Diameter
Number of MCP Stages
Center Hole Diameter
Center Dead Space Diameter
MCP Supply Voltage (In-Out)
MCP-Out to Anode Voltage
Electric Potential at Each
Terminal
Operating Vacuum Condition
Baking Temperature
Baking Time
Baking Vacuum Condition
Gain at 2.0kV
Plate Resistance per MCP
Dark Count at 2.0kV
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
Center hole for beam line
High-speed response
Good output waveform with negligible ringing
Parameter
PRELIMINARY DATA
Ideally suited for reflectron mass spectrometer
MAY 1998
6.7 10
-4
Pa (1 10
MCP-Out: -0.5
MCP-In: -2.5
Anode: GND
10 to 100
-4
Value
1 10
(5 10
69.0
27.0
26.4
12.0
6.0
0.5
12
- *
- *
- *
2
2
5
-6
6
-6
Torr), Ta=+25°C)
) Max.
Pa (Torr)
Pa (Torr)
cps/cm
Unit
mm
mm
mm
mm
mm
M
kV
kV
kV
°C
h
-
-
m
2
MCP ASSEMBLY
WITH CENTER HOLE
FOR REFLECTRON MS
F4294-09
APPLICATION
Reflectron mass spectrometer
Figure 1: Typical Output Waveform
TMCPB0077EA
RESPONSE TIME (0.5ns/div)
RISE TIME: 328ps
FALL TIME: 633ps
FWHM
SUPPLY VOLTAGE (MCP): 1700V
: 592ps
TMCPF0090

Related parts for F4294-09

F4294-09 Summary of contents

Page 1

... PRELIMINARY DATA MAY 1998 Ideally suited for reflectron mass spectrometer The F4294- MCP assembly developed for reflectron mass spectrometers. In spite of having a center hole for beam line, the F4294-09 eliminates ringing phenomena which have been a problem with conventional MCP assemblies with a center hole. ...

Page 2

... MCP ASSEMBLY WITH CENTER HOLE FOR REFLECTRON MS F4294-09 Figure 2: Dimensional Outlines (Unit: mm) MCP DEAD SPACE CENTER HOLE DIA. MCP-OUT LEAD 4 - 3.5 P.C.D. 61 EFFECTIVE AREA P.C.D. 41 HAMAMATSU PHOTONICS K.K., Electron Tube Center 314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193 Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U ...

Related keywords