c10013sk Hamamatsu Photonics, K.K.,, c10013sk Datasheet - Page 3

no-image

c10013sk

Manufacturer Part Number
c10013sk
Description
Flat Panel Sensor High Reliability, High Sensitivity
Manufacturer
Hamamatsu Photonics, K.K.,
Datasheet
Note: “He” is the Hsync count. Phe, Phd and Pg are the Pclk count.
To acquire images in external trigger mode, input an external trigger pulse as shown below. When the time Tvd has passed after
the rising edge of the external trigger pulse, synchronous signals and video signals are output.
Timing chart
To acquire images through an image grabber board, write parameters in the software program or parameter file by referring to
the following timing chart and description.
External trigger mode
Phe
He
Vsync+ (GRABBER)
Hsync+ (GRABBER)
Vsync+ (GRABBER)
Hsync+ (GRABBER)
Pclk+ (GRABBER)
Parameter
Phd
Pg
ExtTrgLemo
ExtTrgGrb
(TTL)
Vsync+
(LVDS)
Effective pixel
Dummy pixel
Effective line
Dummy line
Phe
1st ROW
2nd ROW
Hsync+, Pclk+ and Data 1-12 are the same as internal trigger mode.
· Tmin is defined as 1/Sf (int).
· Tmax is defined as the reciprocal of the minimum value of Sf (ext).
· Tvd=320 µs
Tvd
RECOMMENDATION: 50 % OF FRAME TIME
LAST ROW
1 FRAME
He
From Tmin to Tmax
DUMMY LINES
ENLARGED VIEW
All pulses are counted at the rising edge of Pclk+.
The effective video output is only included in the “Phe” period.
Flat panel sensor
1st ROW
2nd ROW
Pg
Count
1000
1032
1870
1877
1st ROW
56
24
Phe
All pulses are counted in the
“High” period of Hsync+
Phd
KACCC0372EA
2nd ROW
Phe
C10013SK
Phd
KACCC0362EA
3