at9000 Yamatake Corporation, at9000 Datasheet - Page 5

no-image

at9000

Manufacturer Part Number
at9000
Description
Differential Pressure Transmitters
Manufacturer
Yamatake Corporation
Datasheet
Yamatake Corporation
Indicator
The digital LCD indicator (optional) indicates engineering
units and can be set freely between -99999 and 99999 (5
digits). For meter calibration, specify the following items
when placing your order
• Meter calibration range
• Meter calibration unit
• Linear / Square-root for meter indication.
Bolts and nuts materials (for fastening
meter body cover)
Carbon steel (SNB7), 304 SST, 630 SST
Paint
OPTIONAL SPECIFICATIONS
Oil free finish
The transmitter is shipped with oil-free wetted parts.
Adapters for anticorrosion materials
These are adaptor flanges to connect 82 mm pipes made of
anticorrosion materials [excluding ASTM B575 (Equiva-
lent to Hastelloy C-276)] to 54 mm general-purpose pipes.
External zero/span adjustment function
The transmitter can be easily zero/span adjusted in the
field.
Elbow
This is an adaptor for changing the electrical conduit con-
nection port from the horizontal to the vertical direction, if
required by wiring conditions in the field. One or two
elbows may be used as needed.
Conformance to Non SI units
We deliver transmitters set to any Non SI units as speci-
fied.
Safety Transmitter
Select this option to be used as a component of Safety
Instrument System (SIS).
AT9000 is complied with IEC61508, certified according to
Safety Integrity Level2 (SIL-2)
Alarm Output (contact output)
Contact output is prepared as alarm output when alarm
(Output Alarm/Sensor Temp. Alarm) condition is
Various kinds of data can be set using the CommPad or
the HART ® 375 communicator.
Standard
Corrosion-resistant paint (Baked acrylic paint)
Corrosion-proof finish
Corrosion-proof paint (Baked epoxy paint), fungus-
proof finish
Corrosion-resistant finish (silver paint)
Transmitter case is coated with silver paint in addition to
the above corrosion-resistant finish.
- 5 -
detected. It can be set to Normally Open. (When alarm is
detected, Contact ON).
Custom calibration
Calibrate for the specified pressure range at the factory.
PHYSICAL SPECIFICATIONS
Materials
Weight
Approx. 3.6 kg (model GTX30D)
INSTALLATION
Electrical connection
1/2NPT internal thread, M20 internal thread.
Grounding
Resistance 100 Ω max.
Mounting
Can be installed on a 2-inch horizontal or vertical pipe
(can be directly mounted on a process pipe)
Process connection
Rc1/2, 1/2NPT internal thread and Rc1/4, 1/4NPT internal
thread.
• Impact to transmitter can damage sensor module.
Fill fluid
Silicone oil for general purpose models
Fluorine oil for oxygen and chlorine models
Center body
316 SST
Transmitter case
Aluminum alloy
Meter body cover flange
SCS14A (Equivalent to 316 SST) or 316 SST, PVC
For Wetted parts
Adapter flange (option)
SCS14A (Equivalent to 316 SST), PVC
Center body
316 SST (Diaphragm 316L SST)
ASTM B575 (Equivalent to Hastelloy C-276), Tanta-
lum, 316L SST
Vents and plugs
316 SST, PVC
No. SS2-GTX00D-0100 (Rev.8)
WARNING

Related parts for at9000