MIT410-EN Megger, MIT410-EN Datasheet - Page 5

Insulation Testers / Megohmmeters BASSIC CAT3 WITH 50V-1000V PI & DAR

MIT410-EN

Manufacturer Part Number
MIT410-EN
Description
Insulation Testers / Megohmmeters BASSIC CAT3 WITH 50V-1000V PI & DAR
Manufacturer
Megger
Datasheet

Specifications of MIT410-EN

Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Zero offset at probe tips:
Lead resistance zeroing:
Buzzer:
Resistance
Measurement:
Accuracy:
Open circuit voltage:
Short circuit current:
Voltage range
0 to 600 V d.c. ±2% ±2 digits
10 mV to 600 V TRMS sinusoidal (40 to 400 Hz) ±2% ±2 digits
0 to 1000 V on Analog scale
Unspecified input level 0 - 10 mV (40 to 400 Hz)
For non-sinusoidal waveforms additional specification apply:
±3% ±2 digits 101 mV to 600 V TRMS and ±8% ±2 digits 10 mV to 100
mV TRMS
Default Voltmeter:
Frequency:
Capacitance measurement
MIT420, MIT430, MIT481 and MIT485
Measurement range:
Accuracy:
Distance by capacitance:
MIT481 and MIT485
Arithmetic conversion from capacitance measurement on
Default capacitance measurement: 50 nF/m
Capacitance range:
Result storage
Capacity:
Download:
Bluetooth Class:
Range:
0.10Ω typical
Up to 9.00 Ω
Variable limit 1Ω, 2Ω, 5Ω,
10Ω, 20Ω
0.01 kΩ to 1000 kΩ
(0 to 1 MΩ on analog scale)
±3% ±2 digits
5 V ±1 V
20 µA ± 5 µA
Operates at >25 V a.c. or d.c.
on any range except OFF
40 - 450 Hz (40 Hz - 99.9 Hz)
±0.5% ±1 digit (100 Hz to
450 Hz)
100 pF to 10 µF
± 5.0% ±2 digits
40 nF/m to 60 nF/m
>1000 test results
Bluetooth wireless
Class II
up to 10 m
Power Supply:
5 x 1.5 V cells type IEC LR6 (AA, MN1500, HP7, AM3 R6HP) Alkaline
NiMH rechargeable cells may be used.
Battery life: 2200 insulation tests with duty cycle of 5 sec ON /55 sec
OFF @ 1000 V into 1 MΩ
Dimensions
Instrument: 220 x 92 x 50 mm (8.66 in. x 3.63 in. x 1.97 in.)
Instrument and case: 456 x 178 x 89 mm (18 in. x 7 in. x 3.5 in.)
Weight
Instrument only: 590 gms (20.73 oz.), 775 gms (27.22 oz.) with boot
Instrument and case: 1.75kg (3.86 lb)
Fuse
Use only a 500 mA (FF) 1000 V 32 x 6 mm ceramic fuse of high
breaking capacity HBC 50 kA minimum. Glass fuses MUST NOT
be fitted.
Safety Protection
The instruments meet EN 61010-1 (1995) to 600 V phase to earth,
Category IV. Refer to safety warnings supplied.
E.M.C.
In accordance with IEC 61326 including amendment No.1
Temperature effects
Temperature coefficient: <0.1% per °C up to 1 GΩ
Environmental
Operating range:
Operating humidity:
Storage temperature range:
Calibration Temperature:
Maximum altitude:
Dust and water protection:
IP54 Protected against dust and splashing water
The Bluetooth word mark and logo are owned by the Bluetooth SIG, Inc.
and any use of such marks by Megger is under license.
Insulation Resistance and Continuity Testers
-10 to +55° C
90% RH at 40° C max.
-25 to +70° C
+20° C
2000 m
MIT400 Series CAT IV