Specifications of FLUKE-115

Accuracy
+/- 0.1 %
Equipment Type
Digital Multimeters
Display Count
6000
Ranging
Auto, Manual
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
2538790
Notes:
1
2
2 Fluke Corporation
Accuracy specifications
All ac voltage ranges are specified from 1 % to 100 % of range. Because inputs below 1 % of range are not specified,
it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected
from a circuit or are shorted together. For volts, crest factor of ≤ 3 at 4000 counts, decreasing linearly to 1.5 at full
scale. AC volts is ac coupled and ac mV is dc coupled.
Frequency is ac coupled, 5 Hz to 50 kHz for ac voltage. Frequency is dc coupled, 45 Hz to 5 kHz for ac current.
Measurement
DC millivolts
DC volts
Auto volts
AC millivolts
true-rms
AC volts
Continuity
Ohms
Diode test
Capacitance
Lo-Z capacitance
AC amps true-rms
(45 Hz to 500 Hz)
DC amps
Hz (V or A input)
1
true-rms
1
2
115 Multimeter
Range
600.0 mV
6.000 V
60.00 V
600.0 V
600.0 V
600.0 mV
6.000 V
60.00 V
600.0 V
600 Ω
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
40.00 MΩ
2.000 V
1000 nF
10.00 µF
100.0 µF
9999 µF
100 µF to 1000 µF
> 1000 µF
1 nF to 500 µF
6.000 A
10.00 A
20 A overload for 30 seconds max.
6.000 A
10.00 A
20 A overload for 30 seconds max.
99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
Resolution
0.1 mV
0.001 V
0.1 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 Ω
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.001 V
0.01 µF
0.1 µF
1 µF
0.01 A
0.001 A
0.01 A
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.01 V
0.1 V
1 nF
0.001 A
Accuracy ± ([% of reading] + [counts])
0.5 % + 2
0.5 % + 2
2.0 % + 3 (dc, 45 Hz to 500 Hz)
4.0 % + 3 (500 Hz to 1 kHz)
1.0 % + 3 (dc, 45 Hz to 500 Hz)
2.0 % + 3 (500 Hz to 1 kHz)
1.0 % + 3 (45 Hz to 500 Hz)
2.0 % + 3 (500 Hz to 1 kHz)
Beeper on < 20 Ω off > 250 Ω; detects
opens or shorts of 500 µs or longer.
0.9 % + 2
0.9 % + 1
5 % + 2
0.9 % + 2
1.9 % + 2
1.9 % + 2
5 % + 20
10 % + 2 typical
1.5 % + 3
1.0 % + 3
0.1 % + 2
Fluke. Keeping your world
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006-2009 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 2/2009 2634043 D-EN-N Rev C
Modification of this document is not permitted
without written permission from Fluke Corporation.
Ordering
information
Fluke-115
Included
TL75 Test leads, holster, User’s
manual and 9 V battery (installed).
up and running.
Multimeter