... ScopeMeter Series II 190-104 and 190-204: The first high-performance four-channel scopes built for harsh industrial environments Introducing the first high-performance portable oscilloscopes with four independent isolated input channels dust- and drip-proof rating, and a CAT III 1000 V / CAT IV 600 V safety rating. Choose 200 MHz or 100 MHz bandwidth models. Now, plant maintenance engineers and technicians can take a four- channel scope into the harsh world of industrial electronics ...
Page 2
... – all with user selectable scaling of resultant; mathematics A versus B (X-Y-mode); Frequency Spectrum using FFT analysis Acquisition modes Normal, Averaged, Auto, Single Shot, ScopeRecord™ roll, glitch capture, waveform compare with automatic 2 Fluke Corporation 2 Fluke Corporation ScopeMeter 190 Series II ScopeMeter 190 Series II 190C Series 196C, 215C 192C ...
Page 3
... Waveform compare Provides storage and display of a reference waveform for visual comparison with newly acquired waveforms. Reference is derived from an acquired waveform and can be modified in the ScopeMeter or externally using FlukeView Software. Pass/Fail Testing In waveform compare mode, the ScopeMeter can be set up to store only matching (“Pass”) or only non-matching (“Fail”) ...
Page 4
... Accuracy Other meter functions Continuity Diode test Amps Temperature 4 Fluke Corporation ScopeMeter 190 Series II (225C and 215C models only) • AS-i (EN50295, 166 kb/s); • CAN-bus (ISO-11898 Mb/s); • Modbus (EIA-232 up to 115 kb/s and EIA-485 Mb/s); • Foundation Fieldbus H1 (61158 type 1, 31.25 kb/s) ; ...
Page 5
... Up to 64x zoom-out for full record overview 10x zoom-in for maximum detail. Two multiple input TrendPlot records can be saved internally for later recall and analysis. Direct storage on external flash memory drive through USB host port. ScopeMeter 190 Series II Fluke Corporation 5 ...
Page 6
... TrendPlot recording of 4 measurements • On PC, using FlukeView™ Software, or • Direct storage on external flash memory drive through USB host port • On PC, using FlukeView™ Software, or • Internally (in instrument) which can be copied on to external flash memory drive as .BMP-file, through USB host port ...
Page 7
... With the new release V5, the following typenumbers are supported: • Fluke 190C-series (225C, 215C, 199C, 196C, 192C, using an OC4USB or PM9080 interface cable); • Fluke 190B-series (199B, 196B, 192B, using an OC4USB or PM9080 interface cable); • 190-series II (190-204 and 190-104, using USB-cable); • 120-series (123, 124, 125, using an OC4USB or PM9080 interface cable) ...
Page 8
... Probe accessory replacement set for VPS210 Series probes Fluke also offers a wide range of optional accessories like temperature probes, current clamps, high voltage probes, cables, adapters and carrying cases to further assist you in your job. See the Fluke website or contact your distributor for details. ...