F45-1-71-1 Simpson, F45-1-71-1 Datasheet

no-image

F45-1-71-1

Manufacturer Part Number
F45-1-71-1
Description
Digital Panel Meters DCP 4-20mA 12VDC 4.5
Manufacturer
Simpson
Type
DC Process Digital Panel Meterr
Datasheet

Specifications of F45-1-71-1

Equipment Type
DC Process Digital Panel Meter
Display Type
4 - 1/2 Digit 0.56 In
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
• 1/8 DIN Indicator
• 3-1/2 or 4-1/2 digit bright red LED display
• Front panel pops off for easy decimal
• Only 3.12” (79mm) required behind panel
• Optional excitation output
• NEMA 4X enclosure optional
• Din Rail Adapter available
Mounting Requirements
The Falcon series 1/8 DIN indicators require a panel cutout of
1.77" (45mm) high by 3.62" (92mm) wide. To install the Falcon
into a panel cutout, remove the clips from the side of the meter.
Slide the meter through your panel cutout, then slide the mount-
ing clips back on the meter. Press evenly to ensure a proper fit.
Tighten screws.
point setting and display scaling
(page C1)
Dimensions
Specifications
DISPLAY
Type
Height
Decimal Point
Overrange indication
Polarity
POWER REQUIREMENTS
AC Voltages
DC Voltages
Power Consumption
ISOLATION
ACCURACY @25°C
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
Warm-up Time
NOISE REJECTION
NMRR
CMRR
A TO D CONVERSION
Technique
Rate
MECHANICAL
Bezel
Depth
Panel Cutout
Case Material
Weight
Temperature Coefficient
F35 DC (Voltage/Current/Process)
F35 AC/TRMS (Voltage/Current)
F45 AC (Voltage/Current)
F45 DC Current
F45 DC (Voltage/Process)
F45 DC Process/Voltage
F35 DC Process/Voltage
F45 DC Current
F35 DC Current
F45 AC Voltage/Current
F35 AC Voltage/Current
4.04”
3.59”
3.78
1.89
7-segment, red LED
0.56" (14.2mm)
user-programmable,
internally or on the terminal block
most significant digit = "1"; other digits
blank
Automatic, with "-" indication,
"+" indication implied
120 or 220VAC, ±10% 50/60Hz
9-32DCV, ±1%
F35: 3VA, F45: 2VA
250V RMS MAX
±0.02% of reading ± 1 count
±0.1% of reading ± 1 count
±0.05% of reading ±1 count
2A ± 5 counts
±0.1% of reading ± 1 count
2A ± 5 counts, 5A ± 5 counts
±0.5% of reading ± 35 counts
±1% of reading ± 5 counts (45Hz-1KHz)
0 to 55°C
-10 to 60°C
0 to 85% non-condensing
Less than 15 minutes
F35=50dB, 50/60Hz; F45=60dB, 50/60Hz
(w/1K unbalanced @ 60Hz) 90dB min.
Dual slope integration
F35=3 samples per second, nominal;
F45=2.5 samples per second, nominal
3.78" x 1.89" x .51"
2.94"
3.62" X 1.77"
94V-1, UL rated Noryl®
9.0oz (255.1g)
(±0.1% ± 0.5 count)/°C
(±0.01% ± 0.05 count)/°C
(±0.05% ± 0.5 count)/°C
(±0.01% ± 0.1 count)/°C
(±0.05% ± 0.1 count)/°C
.51
1.77”
2.94
5A ± 5 counts
92 mm
3.62”
1.74
more >>
2.28
A
7

F45-1-71-1 Summary of contents

Page 1

... NMRR F35=50dB, 50/60Hz; F45=60dB, 50/60Hz CMRR (w/1K unbalanced @ 60Hz) 90dB min CONVERSION Technique Dual slope integration Rate F35=3 samples per second, nominal; F45=2.5 samples per second, nominal MECHANICAL Bezel 3.78" x 1.89" x .51" Depth 2.94" Panel Cutout 3.62" X 1.77" Case Material 94V-1, UL rated Noryl® ...

Page 2

... Range Resolution Input Impedance Resolution Input Impedance 200mV 100uV >100MEG 2V 1mV >1MEG 20V 10mV >10MEG 200V 100mV >9.9MEG Specifications for F45 Temperature Meters DISPLAY Type Height Decimal Point Overrange indication Polarity POWER REQUIREMENTS AC Voltages DC Voltages Power Consumption ENVIRONMENTAL imped- Operating Temperature ...

Page 3

... DCV Process 80 (F45) J T/C K T/C 81 (F45) S T/C 82 (F45) 83 (F45) T T/C 84 (F45) E T/C R T/C 85 (F45) DCmV 86 (F45) 90 (F45) RTD Pt100 Ohm 91 (F35) 20-199.9 Hz RMS 92 (F35) 20-1999 Hz RMS 93 (F35) 20-199.9 Hz Sq. Wave 94 (F35) 20-1999 Hz Sq. Wave (not available with AC, temperature, or frequency inputs) ...