MTB080009A Elpac, MTB080009A Datasheet

PS 9VDC 7.2AMP INT MED 2X4

MTB080009A

Manufacturer Part Number
MTB080009A
Description
PS 9VDC 7.2AMP INT MED 2X4
Manufacturer
Elpac
Datasheet

Specifications of MTB080009A

Input Voltage
100 VAC to 240 VAC
Product
Switching
Commercial/medical
Medical
Output Power Rating
65 W
Number Of Outputs
1
Output Voltage (channel 1)
9 V
Output Current (channel 1)
7.2 A
Mounting Style
Panel
Dimensions
101.6 mm L x 50.8 mm W x 25.7 mm H
Open Frame/enclosed
Open Frame
Power Supply Type
Switching
Power Supply Output Type
Fixed
No. Of Outputs
1
Output Voltage
9V
Output Current
7.2A
Power Rating
65W
Length
101.6mm
Width
50.8mm
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
A
1-888-357-2280
saleselpac@iccus.com
www.iccus.com
+44.1383.432920
saleseurope@iccus.com
TECHNICAL SUPPOrT
WOrLDWIDE
EUrOPE
Notes
**Ripple and noise measured with 20MHz bandwidth; 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor.
Input Voltage
Input Frequency
Input Current
Inrush Current
Zero Load
Power Consumption
Earth Leakage Current
Patient Leakage Current
Input
*visit www.iccus.com for complete details
5-YEAr LIMITED WArrANTY*
85 – 264VAC
100 – 240VAC Nominal
47 – 63Hz
<2A rms
<37A at 230VAC cold start
<0.3W
<75µA @ 132VAC @ 60Hz
<120µA @ 264VAC @ 60Hz
<40µA @ 132VAC @ 60Hz
<60µA @ 264VAC @ 60Hz
Medical Switch Mode Power Supply
MTB080 Series
Output Voltage
Total Regulation
Minimum Load
Start-Up Delay
Hold-Up Time
Ripple & Noise
Over Voltage Protection
Over Temperature Protection
Over Current Protection
Short Circuit Protection
Output
High Efficiency: Level V
High Power Density 12.5W/in
Output Floating
Lifetime expectation >5 years
Hold-up Time 28ms at full load
Medical Approval - EN60601-1 Class I
See Table
+/-5%
No minimum load required
<0.2s
>28ms at full load
<1% pk-pk **
120 -150%
Active - Recoverable; plus
Passive - Non Recoverable
110 - 190%
shutdown, auto-restart
(hiccup mode)
65 - 100 Watts
3

Related parts for MTB080009A

MTB080009A Summary of contents

Page 1

... TECHNICAL SUPPOrT WOrLDWIDE 1-888-357-2280 saleselpac@iccus.com www.iccus.com EUrOPE +44.1383.432920 saleseurope@iccus.com Input Input Voltage Input Frequency Input Current Inrush Current Zero Load Power Consumption Earth Leakage Current Patient Leakage Current Notes *visit www.iccus.com for complete details **Ripple and noise measured with 20MHz bandwidth; 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor. ...

Page 2

... Model Number Output Voltage MTB080009A 9.0V MTB080012A 12.0V MTB080015A 15.0V MTB080024A 24.0V Notes 1) Maximum load current with natural convection cooling 2) Maximum peak load lasting seconds with natural convection cooling, or maximum continuous output current with minimum 5 CFM airflow. 3) Includes initial setting, line regulation, load, regulation, and thermal drift. ...

Related keywords