FLUKE-1625 Fluke, FLUKE-1625 Datasheet
FLUKE-1625
Specifications of FLUKE-1625
FLUKE-1625 Summary of contents
Page 1
... Adjustable limits - for quicker testing Stakeless testing The Fluke 1623 and 1625 earth ground testers are able to measure earth ground loop resistances using only clamps. With this test method, two clamps are placed around the earth ground rod and each are con- nected to the tester ...
Page 2
... The most complete testers The Fluke 1623 and 1625 are distinctive earth ground testers that can perform all four types of earth ground measurement: • 3- and 4-Pole Fall of Potential (using stakes) • 4-Pole Soil Resistivity testing (using stakes) • Selective testing (using 1 clamp and stakes) • ...
Page 3
... H Monitoring and R with error indicator S H Automatic range selection Measurement is not performed if the current through the current clamp is too low 3 Fluke Corporation 1623 and 1625 GEO Earth Ground Testers Measuring range Intrinsic error 0.001 Ω to 19.99 kΩ ± rdg + > 128 Hz (125 Hz on request) Max 100 kΩ ...
Page 4
... EXT Automatic range selection The information regarding stakeless ground loop measurements is only valid when used in conjunction with the recommended current clamps at the minimum distance specified. 4 Fluke Corporation 1623 and 1625 GEO Earth Ground Testers Measuring range Intrinsic error 0.001 Ω to 19.99 kΩ ...
Page 5
... Fluke Corporation 1623 and 1625 GEO Earth Ground Testers -10 ° ...
Page 6
... Short circuit current 250 mA ac Measuring frequency 94, 105, 111, 128 Hz selected manually or automatic. (AFC function R* Noise rejection 120 dB (16 Max. overload U rms 6 Fluke Corporation 1623 and 1625 GEO Earth Ground Testers socket to socket = 0 V “ to each other in any combination, max Fullwave rectification ...
Page 7
... Automatic check if error is kept within the limits required by IEC61557-5. If after a measurement of probe-, auxiliary earth electrode- and earthing resistance, a measurement error of higher than assumed because of the influencing conditions (see diagram), the display shows a warning symbol 7 Fluke Corporation 1623 and 1625 GEO Earth Ground Testers Designation ...
Page 8
... Additonal error because of probe- and auxiliary earth typ. electrode resistance Measuring error of R and Measuring time Minimal current in single branch to be measured Max. interference current through transformer 8 Fluke Corporation 1623 and 1625 GEO Earth Ground Testers Resolution 1 mΩ 10 mΩ 100 mΩ 1 Ω ...
Page 9
... R K Formation of measured value R display = 0.000 Ω, variable from 0.000 to 29.99 Ω by means of measuring adjustment. * Value of setpoint entry Fluke Corporation 1623 and 1625 GEO Earth Ground Testers Resolution 0.001 Ω ± rdg + 2 digit) 0.01 Ω 0.1 Ω 1 Ω ...
Page 10
... Fluke-1623 Kit Basic GEO Earth Ground Tester Kit Fluke-1623 Basic GEO Earth Ground Tester EI-1623 Selective/Stakeless Clamp Set for 1623 Fluke-1625 Kit Advanced GEO Earth Ground Tester Kit Fluke-1625 Advanced GEO Earth Ground Tester EI-1625 Selective/Stakeless Clamp Set for 1625 Optional accessories ...