AD8220WARMZ-RL Analog Devices Inc, AD8220WARMZ-RL Datasheet - Page 25

no-image

AD8220WARMZ-RL

Manufacturer Part Number
AD8220WARMZ-RL
Description
Auto Open Mkt W/AECQ00
Manufacturer
Analog Devices Inc
Datasheet

Specifications of AD8220WARMZ-RL

Amplifier Type
J-FET
Number Of Circuits
1
Output Type
Rail-to-Rail
Slew Rate
2 V/µs
-3db Bandwidth
1.5MHz
Current - Input Bias
25pA
Voltage - Input Offset
250µV
Current - Supply
750µA
Current - Output / Channel
15mA
Voltage - Supply, Single/dual (±)
4.5 V ~ 36 V, ±2.25 V ~ 18 V
Operating Temperature
-40°C ~ 125°C
Mounting Type
Surface Mount
Package / Case
*
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Gain Bandwidth Product
-
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
ELECTROCARDIOGRAM SIGNAL CONDITIONING
The AD8220 makes an excellent input amplifier for next
generation ECGs. Its small size, high CMRR over frequency,
rail-to-rail output, and JFET inputs are well suited for this
application. Potentials measured on the skin range from 0.2 mV
to 2 mV. The AD8220 solves many of the typical challenges of
measuring these body surface potentials. The high CMRR of
the AD8220 helps reject common-mode signals that come in
the form of line noise or high frequency EMI from equipment
in the operating room. Its rail-to-rail output offers a wide
dynamic range allowing for higher gains than would be possible
using other instrumentation amplifiers. JFET inputs offer a
large input capacitance of 5 pF. A natural RC filter is formed
reducing high frequency noise when series input resistors are
used in front of the AD8220 (see the RF Interference section).
C
A
B
15kΩ
10kΩ
10kΩ
2.2pF
2.2pF
10pF
+5V
–5V
+5V
–5V
24.9kΩ
24.9kΩ
AD8220
OP2177
499kΩ
INSTRUMENTATION
4.12kΩ
OP AMPS
AMPLIFIER
G = +14
+5V
–5V
+5V
–5V
866kΩ
68pF
HIGH-PASS FILTER 0.033Hz
–5V
+5V
OP2177
Figure 67. Example ECG Schematic
12.7kΩ
2.5V
4.7µF
220pF
Rev. B | Page 25 of 28
1.18kΩ
2.5V
1MΩ
G = +50
57.6kΩ
+5V
AD8618
In addition, the AD8220 JFET inputs have ultralow input
bias current and no current noise, making it useful for ECG
applications where there are often large impedances. The MSOP
and the optimal pinout of the AD8220 allow smaller footprints
and more efficient layout, paving the way for next-generation
portable ECGs.
Figure 67 shows an example ECG schematic. Following the
AD8220 is a 0.033 Hz high-pass filter, formed by the 4.7 μF
capacitor and the 1 MΩ resistor, which removes the dc offset
that develops between the electrodes. An additional gain of 50,
provided by the AD8618, makes use of the 0 V to 5 V input
range of the ADC. An active, fifth-order, low-pass Bessel filter
removes signals greater than approximately 160 Hz. An
buffers, inverts, and gains the common-mode voltage taken at
the midpoint of the AD8220 gain setting resistors. This right-
leg drive circuit helps cancel common-mode signals by
inverting the common-mode signal and driving it back into the
body. A 499 kΩ series resistor at the output of the
limits the current driven into the body.
14kΩ
2.5V
LOW-PASS FIFTH ORDER FILTER AT 157Hz
14kΩ
47nF
+5V
AD8618
19.3kΩ
19.3kΩ
33nF
2.5V
AD8618
33nF
4.99kΩ
1.15kΩ
+5V
14.5kΩ
14.5kΩ
68nF
2.5V
AD8618
22nF
+5V
2.7nF
500Ω
OP2177
AD8220
REF
+5V
AD7685
ADC
OP2177
REFERENCE
ADR435
4.7µF

Related parts for AD8220WARMZ-RL