EVAL-AD5382EBZ Analog Devices Inc, EVAL-AD5382EBZ Datasheet - Page 5

no-image

EVAL-AD5382EBZ

Manufacturer Part Number
EVAL-AD5382EBZ
Description
Evaluation Board I.c.
Manufacturer
Analog Devices Inc
Datasheets

Specifications of EVAL-AD5382EBZ

Number Of Dac's
32
Number Of Bits
14
Outputs And Type
32, Single Ended
Sampling Rate (per Second)
125k
Data Interface
DSP, I²C, MICROWIRE™, Parallel, QSPI™, SPI™
Settling Time
8µs
Dac Type
Voltage
Voltage Supply Source
Single
Operating Temperature
-40°C ~ 85°C
Utilized Ic / Part
AD5382
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
Q4391457
EVALUATION BOARD SOFTWARE
SOFTWARE INSTALLATION
The EVAL-AD5380/81/82/83EB evaluation kit includes self-
installing software on CD-ROM. If the setup file does not run
automatically, you can run setup.exe from the CD-ROM.
The evaluation board software is compatible with Windows® 95
to Windows XP. Ensure that the Centronics cable connects the
PC to the EVAL-AD5380/81/82/83EB evaluation board.
1.
2.
The AD5380/81/82/83 Evaluation Software dialog box is
displayed. Figure 2 shows the drop-down menus (File, Printer
Port, Dac, Register, and Help).
LOAD DAC CHANNELS
The Dac menu allows user control of the 32/40 DAC channels,
as well as other functions such as hardware reset of the DAC
channels and clearing the outputs.
To load DAC channels:
1.
2.
3.
4.
On the Analog Devices menu, select AD538x Evaluation
Software.
At the prompt, select the relevant device.
On the Dac menu, select Load/Calibrate Dac Channel.
The Load/Calibrate Dac Outputs dialog box is displayed.
In the Enter Channel box, enter the channel selection.
In the Enter Data (Hex) box, type a DAC code to load to
the selected channel.
Press Enter on your keyboard after entering data.
Figure 2.
Figure 3.
Rev. 0 | Page 5 of 12
5.
CHANNEL CALIBRATION
1.
2.
3.
4.
5.
6.
You can also manually program the gain and offset registers.
1.
2.
Each DAC has two associated data registers, the A Register
and the B Register. Click the Toggle Selected Register for
output button to toggle between the two registers.
Click Calibrate Channel to display the Channel
Calibration dialog box.
Type the reference value in the Enter Vref(+) Value (V)
box.
Load full scale and measure the output voltage on the
relevant channel.
Input this value in the appropriate field in the dialog box.
Load zero scale and again enter the measured value in the
appropriate field.
Click Calibrate to calibrate the device to the required
output voltage.
Click Manually Program Calibration Registers. The
Program m and c Registers dialog box is displayed.
Type values for the gain and offset registers and click the
accompanying buttons to program the registers.
EVAL-AD5380/81/82/83EB
Figure 4.
Figure 5.