FLUKE-87-5 Fluke Electronics, FLUKE-87-5 Datasheet - Page 56

DMM INDUSTRIAL TRUE-RMS W/TEMP

FLUKE-87-5

Manufacturer Part Number
FLUKE-87-5
Description
DMM INDUSTRIAL TRUE-RMS W/TEMP
Manufacturer
Fluke Electronics
Series
8xr
Type
Digital (DMM)r
Datasheets

Specifications of FLUKE-87-5

Includes
Battery, Clips, Temperature Probe, Test Leads
Style
Handheld
Display Digits
4.5
Display Type
LCD, Bar Graph
Display Count
20000
Function
Voltage, Current, Resistance, Capacitance, Temperature, Frequency
Functions, Extra
Continuity, Diode Test
Features
Auto Off, Backlight, Clips, Hold, Min/Max/Ave
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Vendor undefined / RoHS compliant by exemption
Other names
2074974
614-1042
87 V
87-5
87-V
FLUKE 87 V
FLUKE-87-V
Users Manual
48
(45 Hz to 2 kHz)
1.
2.
3.
4.
5.
6.
mA
(45 Hz to 2 kHz)
mA
A B
\
[
AF
Function
AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sine wave input.
AC conversions for Model 87 are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except 400 mA range (5 %
to 100 % of range) and 10 A range (15 % to 100 % or range).
Model 87 is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a
residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit change for readings over 3 % of
range. Using REL to offset this reading may produce a much larger constant error in later measurements
W 10 A continuous up to 35 C; < 20 minutes on, 5 minutes off at 35 C to 55 C. 20 A for 30 seconds maximum; > 10 A unspecified.
Below a reading of 200 counts, add 10 counts.
400 mA continuous; 600 mA for 18 hrs maximum.
60.00 mA
400.0 mA
6.000 A
10.00 A
60.00 mA
400.0 mA
6.000 A
10.00 A
600.0 A
600.0 A
6000 A
6000 A
Range
4
4
6
6
Table 14. Current Function Specifications
0.01 mA
0.1 mA
0.001 A
0.01 A
0.01 mA
0.1 mA
0.001 A
0.01 A
0.1 A
1 A
0.1 A
1 A
Resolution
Model 83
(1.2 % + 2)
(1.2 % + 2)
(1.2 % + 2)
(1.2 % + 2)
(0.4 % + 4)
(0.4 % + 2)
(0.4 % + 4)
(0.4 % + 2)
(1.2 % + 2)
(1.2 % + 2)
(0.4 % + 4)
(0.4 % + 2)
1
5
5
5
5
5
5
Accuracy
Model 87
(1.0 % + 2)
(1.0 % + 2)
(1.0 % + 2)
(1.0 % + 2)
(0.2 % + 4)
(0.2 % + 2)
(0.2 % + 4)
(0.2 % + 2)
(1.0 % + 2)
(1.0 % + 2)
(0.2 % + 4)
(0.2 % + 2)
2, 3
Burden Voltage
1.8 mV/mA
1.8 mV/mA
1.8 mV/mA
1.8 mV/mA
100 V/ A
100 V/ A
100 V/ A
100 V/ A
(typical)
0.03 V/A
0.03 V/A
0.03 V/A
0.03 V/A