FLUKE-789 Fluke Electronics, FLUKE-789 Datasheet - Page 33

PROCESSMETER DMM W/4-20MA SOURCE

FLUKE-789

Manufacturer Part Number
FLUKE-789
Description
PROCESSMETER DMM W/4-20MA SOURCE
Manufacturer
Fluke Electronics
Series
ProcessMeter™r
Type
Digital (DMM)r
Datasheets

Specifications of FLUKE-789

Includes
Battery, Clips, Test Leads
Style
Handheld
Display Digits
4.5
Display Type
LCD, Dual
Display Count
20000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, Diode Test
Features
Auto Off, Backlight, Hold, Min/Max/Ave
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Lead free / RoHS compliant by exemption
Other names
1633984
614-1075
789
Pushbutton
COARSE
COARSE
Table 8. mA Output Adjust Pushbuttons
R
M
h
r
FINE
FINE
X
X
W
W
Adjusts up 0.1 mA
Adjusts up 0.001 mA
Adjusts down 0.001 mA
Adjusts down 0.1 mA
Adjustment
Manually Stepping the mA Output
When the rotary function switch is in the OUTPUT [ mA
position, and the OUTPUT jacks are connected to an
appropriate load, the meter produces a steady mA dc
output. The meter begins sourcing or simulating 0 %. Use
the pushbuttons to step the current up and down in 25 %
increments as shown in Table 9. See Table 10 for mA
values at each 25 % step.
Select either sourcing or simulating by choosing the
SOURCE or SIMULATE output jacks.
If the meter cannot deliver the programmed current
because the load resistance is too high or the loop supply
voltage is too low, dashes (-----) appear on the numeric
display. When the impedance between the SOURCE
jacks is low enough, the meter will resume sourcing.
The COARSE and FINE adjustment
pushbuttons described in Table 8 are available
when manually stepping the mA output.
Using the Current Output Functions
Note
ProcessMeter
25