TPI 293 TPI (Test Products Int), TPI 293 Datasheet - Page 20

DMM CLAMP AUTO-RANGE 4 DIGIT+BAR

TPI 293

Manufacturer Part Number
TPI 293
Description
DMM CLAMP AUTO-RANGE 4 DIGIT+BAR
Manufacturer
TPI (Test Products Int)
Series
293r
Type
Digital (DMM)r
Datasheet

Specifications of TPI 293

Includes
Battery, Case, Test Leads
Style
Clamp
Display Digits
3.75
Display Type
LCD, Bar Graph
Display Count
4000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, Diode Test
Features
Auto Off, Hold, Sleep
Ranging
Auto/Manual
Response
True RMS
Other names
290-1293
293
20
Application Notes (Resistance)
When measuring resistance of a motor, make
sure the power is disconnected prior to testing.
Set up meter following steps under
“Measurement Procedure” on page 20, and
proceed with the following:
Connect the red test lead to one power
Connect the red test lead to the frame of
input line of the motor and the black test
lead to the other power input line of the
motor. In most applications if the reading
is OFL, the motor winding is open.
the motor and the black test lead to the
winding. In most applications if a reading
of 0 Ohms is displayed, the winding is
shorted to the motor frame (ground).