U1731B AGILENT TECHNOLOGIES, U1731B Datasheet - Page 6

Hand-held LCR Meter

U1731B

Manufacturer Part Number
U1731B
Description
Hand-held LCR Meter
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U1731B

Accuracy %
3%
External Width
87mm
External Depth
41mm
No. Of Batteries
1
Lead Free Status / RoHS Status
na
Accuracy is expressed as ± (% of reading + number of least significant digits) at
23 °C ± 5 °C and <75% R.H.
U1732A/U1732B Electrical Specifications
Resistance (Parallel Mode), Test Frequency = 100 Hz/120 Hz
Resistance (Parallel Mode), Test Frequency = 1 kHz/10 kHz
[1] Specifications are based on measurements performed at the test sockets and on battery operation.
[2] DUT and test leads need to be properly shielded by connecting to the guard terminal, if necessary.
Range
10 MW
2000 kW
200 kW
20 kW
2000 W
200 W
20 W
Range
10 MW
2000 kW
200 kW
20 kW
2000 W
200 W
20 W
Maximum Display
9.999 MW
1999.9 kW
199.99 kW
19.999 kW
1999.9 W
199.99 W
19.999 W
Maximum Display
9.999 MW
1999.9 kW
199.99 kW
19.999 kW
1999.9 W
199.99 W
19.999 W
@ 100 Hz
2.0% + 8
0.5% + 5
0.5% + 3
0.5% + 3
0.5% + 3
0.8% + 5
1.2% + 40
@ 1 kHz
2.0% + 8
0.5% + 5
0.5% + 3
0.5% + 3
0.5% + 3
0.8% + 5
1.2% + 40
6
Accuracy
Accuracy
@ 120 Hz
2.0% + 8
0.5% + 5
0.5% + 3
0.5% + 3
0.5% + 3
0.8% + 5
1.2% + 40
@ 10 kHz
3.5% + 10
2.0% + 10
1.5% + 5
1.5% + 5
1.5% + 5
2.0% + 10
2.5% + 200
Note
After open cal.
After open cal.
After short cal.
After short cal.
Note
After open cal.
After open cal.
After short cal.
After short cal.
-
-
-
-
-
-