U2741A AGILENT TECHNOLOGIES, U2741A Datasheet

MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT

U2741A

Manufacturer Part Number
U2741A
Description
MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U2741A

Dmm Type
Bench
Voltage Measuring Range Dc
100 MV - 300 V
Voltage Measuring Range Ac
100 MV - 250 V
Current Measuring Range Dc
10mA To 2A
Current Measuring Range Ac
10mA To 2A
Lead Free Status / RoHS Status
na
Agilent
U2741A USB Modular 5.5 Digits Digital Multimeter
Data Sheet

U2741A Summary of contents

Page 1

... U2741A USB Modular 5.5 Digits Digital Multimeter Agilent Data Sheet ...

Page 2

... Agilent Development Environments (ADEs) Introduction The Agilent U2741A is a 5.5 digits digital multimeter (DMM), which is the latest addition to the Agilent’s USB modular family. It can operate as standalone modular unit when used with the U2781A USB modular instrument chassis. It comes with various features and functions to meet your needs today. ...

Page 3

... Data can be stored for analysis when this option is enabled. The data capture interval range can be set from 100 s. With this feature, you can start the U2741A to take measurement over duration and check your results when the test is completed. Data can be stored to a CSV (comma separated value) fi ...

Page 4

Standard Shipped Accessory • AC/DC adapter • Power cord • Standard test lead • USB Standard-A to Mini-B interface cable • L-Mount kit (used with the modular instrument chassis) • Agilent Automation-Ready CD (contains the Agilent ...

Page 5

General Characteristics Remote interface Power consumption Operating environment Storage compliance Safety compliance Measurement compliance EMC compliance Common mode rejection Normal mode rejection ratio (NMRR) Shock and vibration I/O connector Dimensions (W × D × H) Weight Warranty Calibration [1] Applicable ...

Page 6

Product Specifi cations DC specifi cations [1] Function Range [2] Voltage 100.000 mV 1.00000 V 10.0000 V 100.000 V 300.000 V Current [3] 10.0000 mA 100.000 mA 1.0000 A 2.0000 A 100.000 Ω Resistance [4] 1.00000 kΩ 10.0000 kΩ 100.000 ...

Page 7

AC specifi cations AC accuracy for voltage Function Range [2] Voltage 100.000 mVrms 1.00000 V 10.0000 V 100.000 V [4] 250.000 V [1] Specifi cations are based on 30 minutes warm-up time and calibration temperature within 18 °C to ...

Page 8

Temperature coeffi cient for current (0 º ºC, 28 º ºC) Range Temperature 10.0000 mA, coeffi cient 100.000 mA, 1.00000 A, 2.0000 A [1] Frequency specifi cations Frequency accuracy Function Range Frequency 20 to 300 kHz ...

Page 9

... For more information on repair and calibration services, go to: www.agilent.com/fi nd/removealldoubt www.agilent.com www.agilent.com/fi nd/U2741A For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/fi nd/contactus ...