P01105101Z CHAUVIN ARNOUX, P01105101Z Datasheet - Page 6

no-image

P01105101Z

Manufacturer Part Number
P01105101Z
Description
CLAMP, MINI 01 1000/1 150A
Manufacturer
CHAUVIN ARNOUX
Series
MINIr
Datasheet

Specifications of P01105101Z

Dmm Type
Clamp
Current Range Ac
2A To 150A
Dmm Response Type
True RMS
Dmm Functions
AC Current
Current Ac Max
150A
Ranging
Manual
Lead Free Status / RoHS Status
na
Clamp-on ammeter for AC current
MINI 09 model
Small and compact, the MINI 09 clamp-on ammeter is
the ideal complement for any multimeter to measure AC
currents in low-power tertiary or industrial applications.
Its DC voltage output helps to overcome the low sensitivity
of certain AC measurement instruments.
Output:
Double-insulated cable 1.5 m long,
terminated by 2 insulated elbowed
male banana connectors Ø 4 mm
Bandwidth: 48 Hz … 500 Hz
Clamping capacity:
Cable Ø max 10 mm
Maximum currents:
I < 150 A permanent from 65 Hz … 500 Hz
Influence of temperature:
≤ 0.2 % per 10 K
Influence of adjacent conductor:
≤ 2 mA / A at 50 Hz
Influence of conductor position in jaws:
≤ 0.1 % at 50/60 Hz
Influence of frequency:
≤ 3 % from 65 Hz to 500 Hz
(1) Conditions of reference: 23 °C ± 3 °K, 20 to 75 % RH, sinusoidal signal with frequency of 48 Hz to 500 Hz, distortion factor < 1 % with no DC component, external DC magnetic field
(2) With clamp closed.
(3) Vibrations expressed in mm peak, scanning of 1 octave/minute for 10 minutes on 3 axes.
Non-contractuel document
906131099 - Ed 1 - 04
Calibre
Sensitivity
Calibre
Measurement range
Accuracy in %
Phase shift
Output signal
To order
MINI 09 clamp-on AC ammeter with operating manual
< 40 A/m, no external AC magnetic field, no external conductor with circulating current, conductor centred for measurement, measurement instrument load impedance ≥ 50 kΩ.
Description
Main specifications
Electrical specifications
1 mV
150 A AC
dc
/ A AC
≤ 10 % + 0.2 A
1 A … 5 A
(1)
100 mV
≤ 6 % + 0.2 A
5 A … 15 A
Operating temperature:
-10 °C to +50 °C
Storage temperature:
-40 °C to +80 °C
Relative humidity for operation:
0 to 85% RH decreasing linearly above 35 °C
Operating altitude:
0 to 2,000 m
Casing protection rating (leakproofing):
IP40
Fall height: 1.5 m (IEC 68-2-32)
Shock protection:
100 g / 6 ms / half-period (IEC 68-2-27)
Vibration resistance
5-15 Hz (1.5 mm), 15-25 Hz (1 mm),
25-55 Hz (0.25 mm) (IEC 68-2-6)
Self-extinguishing capability:
casing UL94 V2
Dimensions: 130 x 37 x 25 mm
Weight: approx. 180 g
Colours: black casing
dc
Mechanical specifications
Not specified
/ A AC (15 V
(2)
150 A
(EN 60529 Ed.1995)
≤ 3 % + 0.2 A
15 A … 40 A
dc
1.04
at 150 A)
(3)
(1/1)
:
40 A … 150 A
≤ 4 %
Electrical safety:
Instrument with double insulation or rein-
forced insulation between the primary, the
secondary and the grippable part located
under the guard as per EN 61010-1
Ed. 2:2001, EN 61010-2-031 Ed. 2002 &
EN 61010-2-032 Ed. 2003
- 600 V category III, pollution degree 2
- 300 V category IV, pollution degree 2
Electromagnetic compatibility:
CE-certified equipment compliant with
standard EN61326-1 (Ed. 97) + A1 (Ed. 98)
+ A2 (Ed. 01)
- Emission: stipulations for Class B
- Immunity: stipulations for equipment
equipment (domestic use).
used intermittently on industrial sites.
Safety specifications
P01105109Z
Reference
MINI Series