TPP1000 Tektronix, TPP1000 Datasheet - Page 5

PASSIVE VOLTAGE PROBE, OSCILLOSCOPE

TPP1000

Manufacturer Part Number
TPP1000
Description
PASSIVE VOLTAGE PROBE, OSCILLOSCOPE
Manufacturer
Tektronix
Type
Passive Prober
Datasheet

Specifications of TPP1000

Test Probe Type
Oscilloscope
Test Probe Functions
Voltage
Bandwidth
1GHz
Cable Length
1.3m
Attenuation
10×
Cable, Length
1.3 m
Input Impedance
10 MΩ
For Use With
MSO/DPO4000B & MSO/DPO5000 Series Mixed Signal Oscilloscopes
Lead Free Status / RoHS Status
na
Lead Free Status / RoHS Status
na
Search step 1: You define what you would like to find.
Search step 2: Wave Inspector automatically searches through the record and marks each
event with a solid colored triangle. You can then use the Previous and Next buttons to
jump from one event to the next.
Play/Pause
A dedicated Play/Pause front-panel button scrolls the waveform across the
display automatically while you look for anomalies or an event of interest.
Playback speed and direction are controlled using the intuitive pan control.
Once again, turning the control further makes the waveform scroll faster and
changing direction is as simple as turning the control the other way.
Digital phosphor technology enables greater than 250,000 wfm/s waveform capture rate
and real-time color grading on the MSO/DPO5000 Series.
User Marks
Press the Set Mark front-panel button to place one or more marks on
the waveform. Navigating between marks is as simple as pressing the
Previous (←) and Next (→) buttons on the front panel.
Search Marks
The Search button allows you to automatically search through your long
acquisition looking for user-defined events. All occurrences of the event
are highlighted with search marks and are easily navigated to, using the
front-panel Previous (←) and Next (→) buttons. Search types include
edge, glitch, width, timeout, runt, pattern, state, setup and hold, transition,
and window.
Digital Phosphor Technology
The MSO/DPO5000 Series’ digital phosphor technology provides you with
fast insight into the real operation of your device. Its fast waveform capture
rate – greater than 250,000 wfm/s – gives you a high probability of quickly
seeing the infrequent problems common in digital systems: runt pulses,
glitches, timing issues, and more.
Waveforms are superimposed with one another and waveform points are
color coded by frequency of occurrence. This quickly highlights the events
that occur more often over time or, in the case of infrequent anomalies,
occur less often.
With the MSO/DPO5000 Series, you can choose infinite persistence
or variable persistence, determining how long the previous waveform
acquisitions stay on-screen. This allows you to determine how often an
anomaly is occurring.
Mixed Signal Oscilloscopes — MSO5000, DPO5000 Series
www.tektronix.com
5