PMM511A PMK, PMM511A Datasheet - Page 2

PASSIVE PROBE, 500MHZ 10:1

PMM511A

Manufacturer Part Number
PMM511A
Description
PASSIVE PROBE, 500MHZ 10:1
Manufacturer
PMK
Datasheet

Specifications of PMM511A

Test Probe Type
Oscilloscope
Accuracy
± 2%
Approval Category
300V CATII
Attenuation
10
Capacitance Max
11pF
Frequency Measuring Range
500 MHz
Resistance
RoHS Compliant
For Use With
Oscilloscopes
PMM 511A
Datasheet
This datasheet supersedes all previously published material. Specifications that are defined typical
are not guaranteed and are published as general information to the user. The instrument should
have warmed up for at least 20 minutes and the environmental conditions must not exceed the spe-
cified limits of the probe. Note that specifications are subject to change without notice.
Electrical specifications
Attenuation ratio
Voltage coefficient
System bandwidth
System risetime
Maximum rated input voltage
Measurement category I:
Measurement category II:
Pollution degree
Voltage derating
(1) Connect to oscilloscope with an input impedance of 1 MΩ ± 1 %.
400
300
200
100
0
0,1
Typical voltage derating PMM511A
(1)
Measurement category I
1
Frequency [MHz]
10
100
10:1
0.0025 % / V
500 MHz
0.7 nsec
400 V rms
1250 V transient overvoltage
300 V rms CAT II
2
1000
Note that the max. input voltage
rating of the probe decreases as
the frequency of the applied signal
increases.
© 2007 PMK GmbH
± 2 % at DC
(typical)
(-3 dB)
(10 % – 90 %)
(typical)
2