ES1B 10-70C Omron, ES1B 10-70C Datasheet - Page 4

no-image

ES1B 10-70C

Manufacturer Part Number
ES1B 10-70C
Description
Sensor Hardware & Accessories IR Thermosensor 10-70C
Manufacturer
Omron
Datasheets

Specifications of ES1B 10-70C

Lead Free Status / Rohs Status
Lead free / RoHS Compliant
■ One-point Input Shift
Preparations
• Set a temperature input range that is suitable for the input specifi-
• Prepare a thermometer to measure the temperature of the sensing
Configuration for Offsetting the Infrared
Thermosensor Input (Figure 1)
Example for the E5CN
1. Adjust the temperature of the sensing object (C) in the configura-
2. Check the temperature of the sensing object (C) and the tempera-
3. Check the temperature of the sensing object (C) and the tempera-
Diagram of One-point Input Shift
4
cations of the Infrared Thermosensor.
object as shown in figure 1, below.
Indicated temperature
after offset
(e.g., 120˚C)
Indicated temperature
before offset
(e.g., 110˚C)
tion shown in figure 1 to near the set point. We will assume that
the temperature indicated on the thermometer is the actual tem-
perature of the sensing object.
ture indicated on the Controller (A) and set both the upper-limit
and lower-limit temperature input settings to the following value:
Temperature C (sensing object) - Temperature A (Controller)
ture indicated on the Controller (A) again. If they are about the
same, then the offset has been properly set.
(A) Temperature
Controller
temperature input
C
Lower-limit
insl
Temperature indicated
by Controller (A)
Infrared Thermosensor
10
0
ES1B Infrared
Thermosensor
(+,−)
Output
Power
Supply
temperature input
(C) Sensing
object
C
Upper-limit
insh
(+,−)
Power supply
10
ES1B
Near set point
(e.g., 120˚C)
+
(B) Thermometer
Input offset value
(e.g., 10˚C)
After offset
Before offset
Temperature
indicated by
thermometer (B)
■ Two-point Input Shift
Use a two-point input shift to output more accurate display values.
Preparations
Refer to the preparations for a one-point input shift.
1. The input value is shifted at two points: near room temperature
2. Use the following formulas to calculate the upper-limit tempera-
Diagram of Two-point Input Shift
Lower-limit Temperature Input Setting
Upper-limit Temperature Input Setting
3. Set both the upper-limit and lower-limit temperature input settings
4. Although we have used two points, near room temperature and
Indicated temp. after
offset, X2 (e.g., 110˚C)
Indicated temp. before
offset,Y2 (e.g., 105˚C)
Indicated temp. after
offset, Y1 (e.g., 40˚C)
Indicated temp. after
offset, X1 (e.g., 25˚C)
i n s l
i n s h
and near the set point. Check the temperature of the sensing
object (C) and the temperature indicated on the Controller (A) at
both near room temperature and near the set point.
ture input and lower-limit temperature input settings based on the
values checked above.
and then check the temperature of the sensing object (C) and the
temperature indicated on the Controller (A) both near room tem-
perature and near the set point.
near the set point, accuracy can be increased further by using
another point within the measurement temperature range other
than the set value of room temperature.
Upper limit of
set range, YH
(e.g., 260˚C)
Lower-limit
temperature
input insl
=
=
Temperature indicated
Lower limit of
YL (e.g., 0˚C)
by Controller (A)
set range,
YH−Y1
YL−Y1
Y2−Y1
Y2−Y1
0
X1, Room temp.
×{(X2−Y2)−(X1−Y1)}+(X1−Y1)
×{(X2−Y2)−(X1−Y1)}+(X1−Y1)
(e.g., 25˚C)
X2, Near set point
(e.g., 110˚C)
Upper-limit
temperature input insh
Temperature
indicated by
thermometer (B)
Before offset
After offset

Related parts for ES1B 10-70C