8808A Fluke, 8808A Datasheet - Page 7

no-image

8808A

Manufacturer Part Number
8808A
Description
Digital Multimeter, 5.5 Digit Resolution
Manufacturer
Fluke
Datasheet
6
Fluke Corporation
Frequency
Continuity
Diode Test
8808A
100 mV to
750 V
Range
Notes:
[1]
[2]
Gate Time...................................................................131 ms
Measurement Method .............................................AC-coupled input using the ac voltage measurement function.
Settling Considerations..........................................When measuring frequency after a dc offset voltage change, errors
Measurement Considerations ...............................To minimize measurement errors, shield inputs from external noise
Continuity Threshold ..............................................20
Test Currents ............................................................1 mA
Response Time ........................................................100 samples/sec with audible tone
Rate…….…………………………………………… Fast
Maxiumum Reading………………………………199.99
Resolution……….….. ……………….……………0.01
Response Time .........................................................100 samples/sec with audible tone
Rate………………………………………………….Fast
Maxiumum Reading………………………..….…1.9999 V
Resolution….………..….. ………………………..0.1 mV
[1,2]
Input > 100 mV
Limited to 8* 10
20 kHz – 200 kHz
200 kHz – 1 MHz
2 kHz – 20 kHz
20 Hz – 2 kHz
Frequency
7
V Hz
0.01 + 0.002
0.01 + 0.002
0.01 + 0.002
0.01 + 0.004
23 °C ± 5 °C
90 days
may occur. For the most accurate measurement, wait up to 1 second to
allow input blocking RC time constant to settle.
when measuring low voltage, low frequency signals.
Fluke 8808A Digital Multimeter Extended Specifications
Accuracy
Accuracy
0.01 + 0.003
0.01 + 0.003
0.01 + 0.003
0.01 + 0.006
23 °C ± 5 °C
1 year
Outside 18 – 28 °C
Coefficient/°C
0.002 + 0.001
0.002 + 0.001
0.002 + 0.001
0.002 + 0.002
Temperature
Fluke Corporation 7