35XPA

Manufacturer Part Number
35XPA
Description
Multimeter, Digital; 1000 mV, 10, 100, 1000 VDC Voltage, Range, DC; RS-232; LCD
Manufacturer
Amprobe
Type
DMMr
Datasheet

Specifications of 35XPA

Battery Type
9V
Capacitance
400 uF
Capacitance Range
4 nF, 40 nF, 400 nF, 4 uF, 40 uF, 400 uF, 4 mF
Current, Range
400 uA, 4000 uA, 40 mA, 400 mA, 2A (Both AC and DC)
Dimensions
155 × 72 × 32 mm
Display Type
LCD
Frequency Range
4KHz, 40 KHz, 400 KHz, 1 MHz
Resistance, Range
400 Ω, 4 kΩ, 40 kΩ, 400 kΩ, 4 MΩ, 40 MΩ
Temperature, Operating
1832°F
Voltage, Range, Ac
400 mV, 4 V, 40 V, 400 V, 750 V
Voltage, Range, Dc
400 mV, 4 V, 40 V, 400 V, 1000 V
Weight
400 g
35XP-A Digital Multimeter
General Specifications
Specifications
Voltage
Current
Resistance
Amprobe® Test Tools
Display
Polarity
Overrange
Zero
Low Battery Indication
Auto Power Off
Measurement Rate
Operating Environment
Storage Temperature
Accuracy
Temperature Coefficient
Altitude
Power
Battery Life
Dimensions (H x W x D)
Weight
Warranty
Function
DC Volts
AC Volts (45 Hz to 500 Hz)
Resolution
Input Impedance
Overload Protection
DC Current
AC Current (45 Hz to 500 Hz)
Resolution
Input Protection
Burden Voltage
Resolution
Open Circuit Voltage
Overload Protection
Ranges
Ranges
Ranges
Ranges
3-3/4 digit liquid crystal display (LCD) with a maximum reading of 3999
Automatic, positive implied, negative polarity indication
(OL) or (-OL) is displayed
Automatic
The “M” is displayed when the battery voltage drops below the operating level
Approx. 10 minutes
2 times per second, nominal
0 °C to 50 °C at <70 % R.H.
-20 °C to 60 °C, 0 to 80 % R.H. with battery removed from meter
Stated accuracy at 23 °C ± 5 °C, <75 % relative humidity
0.1 x (specified accuracy) per °C (0 °C to 18 °C, 28 °C to 50 °C)
2000 m
Single standard 9 volt battery, NEDA 1604, JIS 006P, IEC 6F22
150 hours typical with carbon-zinc, 300 hours typical with alkaline
155 mm x 72 mm x 32 mm
Approx. 400 g including battery
One-year
Range
400 mV, 4 V, 40 V, 400 V, 1000 V
400 mV (45 Hz to 100 Hz)
4 V, 40 V, 400 V, 750 V
100 μV
400 mV: >100 MW; 4 V: 10 MW; 40 V to 1000 V: 9.1 MW
1000 V DC or 750 V AC RMS
400 μA, 4000 μA, 40 mA, 400 mA
2 A
400 μA, 4000 μA, 40 mA, 400 mA
2 A
0.1 μA
2 A/1000 V fast blow ceramic fuse 6.3 mm x 32 mm on A input
500 mV max. (2 V max. on 4 mA, 400 mA ranges)
400 W, 4 kW, 40 kW, 400 kW, 4 MW
40 MW
100 mW
-0.45 V DC typical, (-1.2 V DC on 400 W range)
1000 V DC or 750 V AC RMS
± (0.5 % rdg + 1 dgt)
± (1.5 % rdg + 5 dgts)
± (1.5 % rdg + 5 dgts)
± (1.0 % rdg + 2 dgt)
± (2.0 % rdg + 3 dgts)
± (1.5 % rdg + 5 dgts)
± (2.5 % rdg + 5 dgts)
± (1.0 % rdg + 4 dgts)
± (2.0 % rdg + 5 dgts)
Accuracy
www.Amprobe.com
Data Sheet