U1731B Agilent Technologies Test Equipment, U1731B Datasheet - Page 9
![no-image](/images/no-image-200.jpg)
U1731B
Manufacturer Part Number
U1731B
Description
Orange Handheld LCR Meter
Manufacturer
Agilent Technologies Test Equipment
Type
LCR Meterr
Datasheet
1.U1731B.pdf
(11 pages)
Specifications of U1731B
Backlight
Yes
Capacitance Accuracy
Up to 0.7% + 5 lsd
Capacitance Range
0.1 pF to 199.99 mF
Color, Body
Orange
Display Type
Dual-Display LCD
Includes
Quick Start Guide, Alligator clip leads, 9 V Alkaline battery, Certifi cation of Calibration (CoC)
Interface
IR to USB
Resistance Range
0.001 Ω to 9.999 MΩ
Resolution
20,000 Counts
General Specifications
Parameter
Measurements
Tolerance mode
Test frequency (Accuracy =
±0.1% of actual test frequency)
Measuring circuit mode
Display
Backlight
Ranging mode
Test signal level
Measurement rate
Response time
Auto power-off
Power supply
Power consumption
Input protection fuse
Battery life
Low battery indicator
Operating environment
Storage environment
Temperature coefficient
Weight
Dimensions (H x W x D)
Safety and EMC compliance
Calibration
Warranty
U1731A/U1731B
L/C/R/D/Q
1%, 5%, 10%
Test frequency
setting
120 Hz
1 kHz
Inductance (L): Defaults to series mode
Capacitance/Resistance (C/R): Defaults to parallel mode
L/C/R : Maximum display 19999
D/Q: Maximum display 999 (Auto range)
Available for model U1732A/U1732B
Auto and Manual
~0.6 V
1 reading/s, nominal
~1 s/DUT (manual range)
~5 mins without operation
0.1 A/250 V
7 hours (typical) without backlight and based on new alkaline
0 °C to 40 °C; 0 to 70% relative humidity (R.H.)
–20 °C to 50 °C; 0 to 80% R.H. non-condensing
0.15 x (specified accuracy)/°C (0 °C to 18 °C or 28 °C to 40 °C)
330 g
184 mm x 87 mm x 41 mm
IEC 61010-1:2001/EN 61010-1:2001 (2
EN 61326-2-1:2006, ICES-001:2004, AS/NZS CISPR11:2004
One-year calibration cycle recommended
3 years for main unit
3 months for standard shipped accessories
9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
AC power adapter and cord available as options
~40 mA (on battery operation)
0.08 mA after auto power-off
RMS
will appear when the voltage drops below ~ 6.8 V
Actual test
frequency
120 Hz
1010 Hz
9
nd
Edition) Pollution Degree 2, IEC 61326-2-1:2005/
U1732A/U1732B
L/C/R/D/Q/q
1%, 5%, 10%, 20%
Test frequency
setting
100 Hz
120 Hz
1 kHz
10 kHz
Actual test
frequency
100 Hz
120 Hz
1010 Hz
9.6 kHz