GPM80-24 SL Power ( Ault / Condor ), GPM80-24 Datasheet - Page 2

no-image

GPM80-24

Manufacturer Part Number
GPM80-24
Description
Power Supply; 80 W; Single; 24 V; 3.4 A (Max.); 85 to 264 VAC; 2.3 A (Max.);
Manufacturer
SL Power ( Ault / Condor )
Type
Medicalr
Datasheet

Specifications of GPM80-24

Brand/series
GPM80 Series
Configuration
Open-Frame
Current, Input
2.3 A (Max.)
Current, Inrush
Inrush at 240 Vac under cold start conditions will not exceed 60 A.
Current, Leakage
35 μA @ 254 VAC
Current, Output
3.4 A (Max.)
Dimensions
7 in. L x 4.26 in. D x 1.86 in. (Max.) H
Hold Time
20 ms minimum from loss of ac input at full load, nominal line (115 Vac).
Mounting Type
Panel
Noise, Output
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential Mode
Number Of Outputs
1
Operation
Switching
Output
24 VDC @ 3.4 A
Output Type
Single
Overvoltage Protection
124% ± 12% typical.
Power Rating
80 W
Power, Output
80 W
Primary Type
AC-DC
Regulation, Total
2
Special Features
Remote Sense
Standards
UL Listed, CSA Certified, CE Marked
Voltage, Input
85 to 264 VAC
Voltage, Output
24 VDC
* Add “G” suffix to part number for RoHS compliant model. Contact factory for availability.
A. Total regulation is defined as the maximum deviation from the nominal voltage for all steady-state conditions of initial voltage setting, input line voltage and output load.
B. Ratings for unrestricted natural convection cooling; output 1 & 2 combined load not to exceed 14A continuous; total power = 80W.
C. Ratings with 26 cfm forced air cooling; output 1 & 2 combined load not to exceed 16A continuous; total power = 110W.
D. To maintain these regulation conditions, the +5V current must be at least 1/4 of V2 and not greater than 5 times the V2 current. Requires +5V to be adjusted within ±1%
with at least a 1A load to maintain regulation on this input.
E. For Power Fail option, add -PF after the model number.
GPC80/GPM80 MECHANICAL SPECIFICATIONS
Commercial
Model
GPC80A
GPC80B
GPC80C
GPC80D
GPC80 E
GPC80P
GPC80-5
GPC80-12
GPC80-15
GPC80-24
GPC80-28
GPC80-48
ENVIRONMENTAL SPECIFICATIONS
Temperature (A)
Humidity (A)
Shock (B)
Altitude
Vibration (C)
Medical
Model
GPM80A
GPM80B
GPM80C
GPM80D
GPM80E
GPM80P
GPM80-5
GPM80-12
GPM80-15
GPM80-24
GPM80-28
GPM80-48
Suffix*
RoHS
G
G
G
G
G
G
G
G
G
G
G
G
OPERATING
See individual specs
0 to 95% RH
20 g
-500 to 10,000 ft
1.5 g
pk
Output No.
rms’
0.003 g
1
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
1
1
1
1
1
1
2
/Hz
Output
+12 V
+12 V
+12 V
+12 V
+15 V
+24 V
+12 V
+24 V
+15 V
+24 V
+12 V
-12 V
-12 V
-15 V
-12 V
-15 V
-12 V
+5 V
+5 V
+5 V
+5 V
+5 V
+5 V
NON-OPERATING
-40 to +85°C
0 to 95% RH
40 g
-500 to 40,000 ft
5 g
12 V
15 V
24 V
28 V
48 V
-5 V
5 V
rms’
pk
0.026 g
Minimum
Output
1.0 A
1.0 A
0.5 A
2
0 A
0 A
0 A
0 A
0 A
0 A
1 A
0 A
0 A
0 A
1 A
0 A
0 A
0 A
1 A
0 A
0 A
0 A
1 A
0 A
0 A
0 A
0 A
0 A
0 A
0 A
0 A
/Hz
Maximum (B)
A. Units should be allowed to warm up/operate under non-condensing
conditions before application of power.
B. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3
orthogonal axes, total 6 shocks.
C. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to
2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and 1
hr./axis non-operating.
Output
3.5 A
6.7 A
5.3 A
3.4 A
2.9 A
1.7 A
12 A
12 A
12 A
12 A
12 A
12 A
16 A
3 A
1 A
1 A
3 A
1 A
1 A
3 A
1 A
1 A
2 A
1 A
1 A
2 A
1 A
1 A
1 A
2 A
Maximum ( C ) Output Peak
Output
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
4.5 A
1.2 A
9.2 A
7.3 A
4.6 A
3.9 A
2.3 A
12 A
12 A
12 A
12 A
12 A
12 A
20 A
4 A
4 A
4 A
3 A
3 A
2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
1.2 A
4.5 A
1.2 A
2.5 A
9.2 A
7.3 A
4.6 A
3.9 A
2.3 A
16 A
16 A
16 A
16 A
16 A
16 A
22 A
5 A
5 A
5 A
4 A
4 A
Noise P-P
120 mV
120 mV
120 mV
120 mV
120 mV
120 mV
150 mV
150 mV
240 mV
120 mV
120 mV
240 mV
150 mV
150 mV
400 mV
120 mV
120 mV
120 mV
150 mV
240 mV
280 mV
480 mV
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
Regulation (A)
+10%/-5% D
Total
2%
2%
3%
3%
2%
2%
3%
3%
2%
2%
3%
3%
2%
2%
3%
3%
2%
2%
3%
3%
2%
3%
3%
2%
2%
2%
2%
2%
2%