U2000H Agilent Technologies Test Equipment, U2000H Datasheet

no-image

U2000H

Manufacturer Part Number
U2000H
Description
USB POWER SENSOR, 10 MHZ TO 18 GHZ
Manufacturer
Agilent Technologies Test Equipment
Datasheet
Agilent
U2000 Series USB Power Sensors
Data Sheet

U2000H Summary of contents

Page 1

Agilent U2000 Series USB Power Sensors Data Sheet ...

Page 2

Features Perform power measurement without a power meter Frequency range from 9 kHz to 24 GHz (sensor dependent) Dynamic range from 60 dBm to – +20 dBm Internal zeroing capability and external calibration-free measurements Simplifi ed measurement setup with built-in ...

Page 3

Why Zeroing Is Important For U2000 Series USB Power Sensor? Zeroing a power sensor is performed to reduce zero offset and noise impact to improve RF power measurement accuracy. The U2000 Series USB power sensor has two types of zeroing, ...

Page 4

Typical SWR Chart (25 °C ±10 °C) Figure 2: Typical SWR chart for U2000A, 10 MHz to 18 GHz (25 °C ±10 °C) Figure 3: Typical SWR chart for U2001A, 10 MHz to 6 GHz (25 °C ±10 °C) 4 ...

Page 5

Figure 4: Typical SWR chart for U2002A, 50 MHz to 24 GHz (25 °C ±10 °C) Figure 5: Typical SWR chart for U2001A, 9 kHz to 6 GHz (25 °C ±10 °C) 5 ...

Page 6

Power Linearity Table 3: Power Linearity (after zero at ambient environment conditions) Sensor Power U2000A, U2001A, –60 dBm to +20 dBm U2002A, U2004A Power range –60 dBm to –35 dBm –38 dBm to –15 dBm –20 dBm to –9 dBm ...

Page 7

Switching Point The U2000 Series USB power sensors have two paths, a low power path covering –60 dBm to –10 dBm, and a high power path covering –10 dBm to +20 dBm. The power meter automatically selects the appropriate power ...

Page 8

Settling Time Settling Time for Normal Mode and x2 Mode Number of Averages 1 2 Settling Time1 (s) 0.045 0.09 (Normal Mode) Settling Time1 (s) 0.042 0.05 (x2 Mode) Noise Multiplier for Normal Mode and x2 Mode Number of Averages ...

Page 9

Calibration Factor (CF) and Refl ection Coeffi cient (Rho) Calibration Factor (CF) and Refl ection Coeffi cient (Rho) data is provided in the Certifi cate of Calibration (CoC) that comes with the purchase of U2000 Series USB power sensors. This ...

Page 10

Table 7: Calibration factor uncertainties for U2002A Frequency –60 dBm to –10 dBm ± 1.98% 50 MHz to 2 GHz 2 GHz to 14 GHz ± 2.27% 14 GHz to 16 GHz ± 2.34% 16 GHz to 18 GHz ± ...

Page 11

General Characteristics This instrument is designed for indoor use only. The table below shows the general requirements for the U2000 Series USB power sensors. Temperature 0 °C to +55 °C (operating) –30 °C to +70 °C (non-operating) Relative Humidity Operating ...

Page 12

Features Enhanced visualization with larger and more fl exible display formats Powerful graph functions Display multi-channel power measurements (with more than 10 displays on one window) Convenient data logging and storing up to seven days Time-saving options to save and ...

Page 13

Agilent Email Updates www.agilent.com/fi nd/emailupdates Get the latest information on the products and applications you select. Agilent Direct www.agilent.com/fi nd/agilentdirect Quickly choose and use your test equipment solutions with confi dence. Agilent Open www.agilent.com/fi nd/open Agilent Open simplifi es the ...