U1701A Agilent Technologies Test Equipment, U1701A Datasheet - Page 3

no-image

U1701A

Manufacturer Part Number
U1701A
Description
Handheld Capacitance Meter
Manufacturer
Agilent Technologies Test Equipment
Type
LCR Meterr
Datasheet

Specifications of U1701A

Backlight
Yes
Capacitance Accuracy
Up to 0.5% + 3 lsd
Capacitance Range
0.1 pF to 199.99 mF
Color, Body
Orange
Display Type
Dual-Display LCD
Includes
Quick Start Guide, Alligator clip leads, 9 V Alkaline battery, Certifi cation of Calibration (CoC)
Interface
IR to USB
Resistance Range
0.001 Ω to 9.999 MΩ
Resolution
11,000 Counts
Tolerance
1, 5, 10, 20
Wide Range
0.1 pF to 199.99 mF
Tolerance Mode
1%, 5%, 10% and 20%
U1701A/U1701B Electrical Specifications
General Specifications
Accuracy is given as ± (% of output + counts of least significant digit) at 23 °C
± 5 °C, with relative humidity less than 80% R.H.
For example, 1% ±10 = 1% of reading + 10 counts of least significant digit
Capacitance
*Accuracy is specified to measure film capacitor or better. Use Relative mode to zero residual.
Range
1000.0 pF
10.000 nF
100.00 nF
1000.0 nF
10.000 μF
100.00 μF
1000.0 μF
10.000 mF
199.99 mF
Parameter
Display
Power supply
Power consumption
Battery life
Operating temperature
Operating humidity
Altitude
Storage temperature
Storage humidity
Temperature coefficient
Low battery indicator
Weight
Dimensions (H x W x D)
Safety and EMC compliance
Calibration
Warranty
Resolution
0.001 mF
0.001 nF
0.001μF
0.01 mF
0.01 nF
0.01 μF
0.1 pF
0.1 nF
0.1 μF
Accuracy*
0.5% + 3
1% + 10
1% + 5
1% + 5
2% + 5
U1701A/U1701B
4½-digit liquid crystal display (LCD) with a maximum resolution of
11,000 counts and automatic polarity indication
5.6 mA (on battery operation)
~80 hours without backlight and based on new alkaline
Full accuracy at 0 °C to 50 °C
Up to 80% relative humidity (R.H.) for temperatures up to 31 °C,
decreasing linearly to 50% R.H. at 50 °C
0 to 2000 m
0 to 80% R.H. non-condensing
0.1 x (specified accuracy)/ °C (from 0 °C to 18 °C or 28 °C to 50 °C)
320 g
184 mm x 87 mm x 41 mm
IEC 61010-1:2001/EN 61010-1:2001 (2
EN 61326-2-1:2006, ICES-001:2004, AS/NZS CISPR11:2004
One-year calibration cycle recommended
–20 °C to 60 °C
9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
AC power adapter and cord available as options
3 years for U1701A/U1701B
3 months for standard shipped accessories
will appear when the voltage drops below ~ 6.0 V
Measuring rate as full scale
0.006 times/s
0.86 times/s
0.13 times/s
(approx.)
5 times/s
5 times/s
5 times/s
5 times/s
5 times/s
5 times/s
3
nd
Edition) Pollution Degree 2, IEC 61326-2-1:2005/