GPFM115-28 SL Power ( Ault / Condor ), GPFM115-28 Datasheet - Page 2

no-image

GPFM115-28

Manufacturer Part Number
GPFM115-28
Description
Power Supply, Single Output; 115 W; Single; 28 V; 2.9 A (Max.); 85 to 264 VAC;
Manufacturer
SL Power ( Ault / Condor )
Type
Medicalr
Datasheet

Specifications of GPFM115-28

Brand/series
GPFM115 Series
Configuration
Open-Frame
Current, Input
1.7 A (Max.) @ 90 VAC, 60 Hz
Current, Leakage
90 uA @ 132 VAC, 60 Hz (Normal Condition), 250 uA @ 264 VAC (Single Fault Condition)
Current, Output
2.9 A (Max.)
Dimensions
5.25 in. L x 3.3 in. W x 1.5 in. H
Hold Time
25 ms minimum from loss of ac input at full load.
Mounting Type
Panel
Noise, Output
0.5% rms, 1% pk-pk, 20 MHz Bandwidth, differential mode. Measured with noise probe directly across output terminals of the power supply.
Number Of Outputs
1
Operation
Switching
Output
28 VDC @ 4.1 A
Output Type
Single
Power Rating
115 W
Power, Output
115 W
Primary Type
AC-DC
Regulation, Total
2
Special Features
PFC, Remote Sense
Standards
CE Marked to LVD
Voltage, Input
85 to 264 VAC
Voltage, Output
18 VDC
* Add “G” suffix to part number for RoHS compliant model. Contact factory for availability.
A. Maximum continuous current rating for unrestricted convection cooling.
B. Maximum continuous current rating with 150 LFM air or cover option.
C. Add “-C” suffix for cover with fan option.
Efficiency
74 to 81% depending upon model.
Temperature Coefficient
0.03%/°C typical on all outputs.
Fan Output
A n additional 12 VDC (nominal 10.5 V), 100 mA output suitable for power-
EMI/EMC Compliance
A ll models include built-in EMI filtering to meet the following emissions
EMI SPECIFICATIONS
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Conducted RF Immunity
Voltage dip Immunity
Line Frequency Harmonics
Medical Safety Standards
A ll models are Certified to be in compliance with the applicable require-
GPFM115-5
GPFM115-12
GPFM115-13
GPFM115-15
GPFM115-24
GPFM115-28
GPFM115-48
GPFM115 Medical
115 Watt Global Performance Switchers
Notes
General Specifications (continued from p. 1)
ing a DC fan is included in all models. The fan output is referenced to +out.
Due to high pk-pk noise and low regulation, it is not recommended that this
output be used for any other purpose.
requirements:
ments of UL 2601, CSA 22.2 No. 601.1-M90, IEC 601-1 (1988), EN 60601-1:
1990. (Consult factory for approval status).
Medical
Model
Suffix*
RoHS
G
G
G
G
G
G
G
Voltage
Output
5.1 V
12 V
13 V
15 V
24 V
28 V
48 V
Adjustment
COMPLIANCE LEVEL
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-3, 3 V/meter
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-5, 1 kV diff., 2 kV com.
EN61000-4-6, 3V
EN61000-4-11
EN61000-3-2 Class A
Output
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Maximum (A)
Output
12 A
6.7 A
6.2 A
5.3 A
3.3 A
2.9 A
1.7 A
Maximum
20 A
9.6 A
8.8 A
7.7 A
4.8 A
4.1 A
2.4 A
(B)
Temperature (A)
Humidity (A)
Shock (B)
Altitude
Vibration (C)
A . Units should be allowed to warm up/operate under non-condensing
B . Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction,
C . Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to
Environmental Specifications
conditions before application of power. Derate output current and total
output power by 2.5% per °C above 50°C. For operation in a confined
space, moving air may be required. Under all conditions, the cooling vs.
load profile should be such that chassis temperatures do not exceed
90°C for extended periods.
3 orthogonal axes, total 6 shocks.
2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and
1 hr./axis non-operating.
Output
Peak
25 A
10.8 A
9.9 A
8.7 A
5.4 A
4.6 A
2.7A
Regulation
Total
OPERATING
0 to 50°
0 to 95% RH
20 g
-500 to 10,000 ft
1.5 g
2%
2%
2%
2%
2%
2%
2%
pk
rms
. 0.003 g
6.2 ± 0.6 V
14 ± 1.1 V
15 ± 1.1 V
18.5 ± 1.5 V
28 ± 2.5 V
34 ± 2.8 V
55 ± 4.0 V
Setpoint
2
/Hz
OVP
NON-OPERATING
-40 to +85°C
0 to 95% RH
40 g
-500 to 40,000 ft
5 g
rms
Ripple/
pk
Noise
. 0.026 g
1%
1%
1%
1%
1%
1%
1%
2
/Hz
Notes
C
C
C
C
C
C
C
2 of 3