U2000A Agilent Technologies Test Equipment, U2000A Datasheet - Page 6

no-image

U2000A

Manufacturer Part Number
U2000A
Description
USB Power Sensor; 10 MHz - 3 GHz
Manufacturer
Agilent Technologies Test Equipment
Datasheet
Power Linearity
Table 3: Power Linearity (after zero at ambient environment conditions)
Figure 6: Typical U2000 Series USB power sensors Power Linearity at 25 °C,
1
U2000A, U2001A,
U2002A, U2004A
Measurement uncertainty and Linearity specifi cations stated above are applicable for both
External Zeroing and Internal Zeroing (with power ranging from -30 dBm to + 20 dBm).
Sensor
–60 dBm to –35 dBm
–38 dBm to –15 dBm
after zero with associated measurement uncertainty
–20 dBm to –9 dBm
–11 dBm to –5 dBm
–7 dBm to 15 dBm
10 dBm to 20 dBm
Power range
6
–60 dBm to +20 dBm
Power
(25 °C±10 °C)
Linearity
Measurement uncertainty
±3.0%
±1.52%
±1.52%
±1.69%
±1.69%
±1.96%
±1.61%
Linearity
(0-55 °C)
±3.5%
1
1