Specifications of MSOX3024A

Bandwidth
200 MHz
Configuration
Benchtop
Dimensions
14.98"(W) × 8.05"(H) × 5.57"(D)
Display
8.5" WVGA
Memory
100 Kpts
Number Of Channels
4 Analog + 16 Digital
Sample Rate
1 GS/s
Weight
8.5 Lbs.
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characteristics
* Denotes warranted specifications, all others are typical.
Acquisition modes
Characteristics
Normal
Peak detect
Averaging
High Resolution Mode
Segmented (optional)
Waveform measurements
Characteristics
Cursors
Automatic measurements
Counter
Waveform math
Characteristics
Arithmetic
FFT
Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
All 3000 X-Series models
Capture glitches as narrow as 250-ps at all timebase settings
Selectable from 2, 4, 8, 16, 64, ... to 65,536
12 bits of resolution when ≥10 µs/div at 4 GSa/s or ≥20-µs/div at 2 GSa/s
Re-arm time = 1 µs (minimum time between trigger events)
All 3000 X-Series models
• Single cursor accuracy:
• Dual cursor accuracy:
Measurements continuously updated with statistics. Cursors track last selected
measurement. Select up to four measurements from the list below:
• Voltage: peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, pre-shoot,
• Time: period, frequency, counter, + width, - width, burst width, duty cycle, rise time, fall time,
• Count: positive pulse count, negative pulse count, rising edge count, falling edge count
• Mixed: area- N cycles, area- full screen
Built-in frequency counter:
• Source: on any analog or digital channel
• Resolution: 5 digits
• Maximum frequency: bandwidth of scope
All 3000 X-Series models
f (g(t))
g(t): { 1, 2, 3, 4, 1-2, 1+2, 1x2, 3-4, 3+4, 3x4}
f(t): { 1-2, 1+2, 1x2, 3-4, 3+4, 3x4, FFT(g(t)), differentiate d/dt g(t), integrate ∫ g(t) dt, square
root √g(t) }
Where 1,2,3,4 represent analog input channels 1, 2, 3, and 4
Note: Channels 3 and 4 only available on MSO/DSOX3xx4A models
Up to 4Mpts resolution
Set FFT Window to: Hanning, Flat Top, Rectangular, Blackman-Harris
+[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale]
+[DC vertical gain accuracy + 0.5% full scale]*
average- N cycles, average- full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N
cycles, AC RMS- full screen (standard deviation), ratio (RMS1/RMS2)
delay, phase, X at min Y, X at Max Y
21