N9010A-507 Agilent Technologies Test Equipment, N9010A-507 Datasheet - Page 8

no-image

N9010A-507

Manufacturer Part Number
N9010A-507
Description
EXA SIGNAL ANALYZER, 7 GHZ
Manufacturer
Agilent Technologies Test Equipment
Type
Networkr
Datasheet

Specifications of N9010A-507

Bandwidth
10 MHz
Capacity
40 GB (Nominal Internal)
Dimensions
7.0 × 16.8 × 14.5 (177 × 426 × 368) In. (mm)
Display Size
8.4 (213)
Frequency
7 GHz
Gain
+20 dB (Nominal)
Power Consumption
390 W
Resolution, Display
1024 × 768 XGA
Storage Temperature
-40 to 65 °C
Temperature, Operating
5 to 50 °C
Lead Free Status / Rohs Status
RoHS Exempt Product
Amplitude Accuracy and Range Specifications
Frequency response (10 dB input attenuation, 20 to 30 °C, preselector centering applied, σ = nominal standard deviation)
Preamp on (Option P03)
(0 dB attenuation)
Input attenuation switching uncertainty
Total absolute amplitude accuracy (10 dB attenuation, 20 to 30 °C, 1 Hz ≤ RBW ≤ 1 MHz, input signal –10 to –50 dBm, all settings
auto-coupled except Auto Swp Time = Accy, any reference level, any scale, σ = nominal standard deviation)
Preamp on (Option P03)
Input voltage standing wave ratio (VSWR) (≥ 10 dB input attenuation)
Preamp on (Option P03)
(0 dB attenuation)
8
9 kHz to 10 MHz
10 MHz to 3.6 GHz
3.5 to 7.0 GHz
6.9 to 13.6 GHz
13.5 to 22.0 GHz
22.0 to 26.5 GHz
100 kHz to 3.6 GHz
50 MHz (reference frequency)
attenuation > 2 dB , preamp off
At 50 MHz
At all frequencies
9 kHz to 3.6 GHz
100 kHz to 3.6 GHz
10 MHz to 3.6 GHz
3.6 to 7.0 GHz
7.0 to 13.6 GHz
13.6 to 26.5 GHz
10 MHz to 3.6 GHz
9 kHz to 3.6 GHz
3.5 to 7.0 GHz
6.9 to 13.6 GHz
13.5 to 26.5 GHz
Specifi cation
±0.8 dB
±0.6 dB
±2.0 dB
±2.5 dB
±3.0 dB
±3.2 dB
±0.20 dB
±0.40 dB
±(0.40 dB + frequency response)
±0.27 dB (95th Percentile ≈ 2σ)
±(0.39 dB + frequency response)
< 1.2:1 nominal
< 1.5:1 nominal
< 1.6:1 nominal
< 1.9:1 nominal
< 1.7:1 nominal
(continued)
95
±0.4 dB
±0.3 dB
±0.28 dB
±0.08 dB typical
±0.3 dB nominal
±0.5 dB nominal
±0.7 dB nominal
±0.7 dB nominal
th