MSO7054B Agilent Technologies Test Equipment, MSO7054B Datasheet - Page 19

no-image

MSO7054B

Manufacturer Part Number
MSO7054B
Description
Oscilloscope, Mixed Signal, 4-Channel, 500 MHz
Manufacturer
Agilent Technologies Test Equipment
Type
Mixed Signalr
Datasheet

Specifications of MSO7054B

Bandwidth
500 MHz
Brand/series
MSO7000 Series
Configuration
Bench/Portable
Memory
8 Mpts
Number Of Channels
4 Analog + 16 Digital
Sample Rate
4 GS/s
Performance characteristics
Horizontal
Range
Resolution
Time scale accuracy*
Vernier
Delay range
Analog delta-t accuracy
Logic delta-t accuracy
Modes
XY
Reference positions
Segmented memory re-arm time
Trigger system
Sources
Modes
Holdoff time
Trigger jitter
* Denotes warranted specification. Specifications are valid after a 30 minute warm-up period and ±10 °C from firmware calibration procedure.
MSO/DSO701xA: 2 nsec/div to 50 sec/div
MSO/DSO703xA: 2 nsec/div to 50 sec/div
MSO/DSO705xA: 1 nsec/div to 50 sec/div
MSO/DSO710xA: 500 psec/div to 50 sec/div
2.5 ps
≤ ± (15+2* (instrument age in years)) ppm
1-2-5 increments when off, ~25 minor increments between major settings when on
Pre-trigger (negative delay): Greater of 1 screen width or 1 ms
Post-trigger (positive delay): 1 s to 500 seconds
Same channel: ±0.0015% reading ±0.1% screen width ±20 ps
Channel-to-channel: ±0.0015% reading ±0.1% screen width ±40 ps
Same channel example (MSO/DSO705xA):
For signal with pulse width of 10 µs, scope set to 5 µs/div (50 µs screen width),
delta-t accuracy = ±{0.0015% (10 µs) + 0.1% (50 µs) + 20 ps} = 50.17 ns
Same channel: ±0.005% reading ±0.1% screen width ±(1 logic sample period, 1 ns)
Channel-to-channel:
±0.005% reading ±0.1% screen width ±(1 logic sample period) ±chan-to-chan skew
Same channel example:
For signal with pulse width of 10 µs, scope set to 5 µs/div (50 µs screen width),
delta-t accuracy = ±{0.005% (10 µs) + 0.1% (50 µs) + 1 ns} = 51.5 ns
Main, zoom, roll, XY, segmented (optional)
Bandwidth: Max bandwidth
Phase error at 1 MHz: < 0.5 degrees
Z Blanking: 1.4 V blanks trace (use external trigger on MSO/DSO7xx2A,
channel 4 on MSO/DSO7xx4A)
Left, center, right
8 µs (minimum time between trigger events)
MSO7xx2A: Ch 1, 2, line, ext, D15 - D0
DSO7xx2A: Ch 1, 2, line, ext
MSO7xx4A: Ch 1, 2, 3, 4, line, ext, D15 - D0
DSO7xx4A: Ch 1, 2, 3, 4, line, ext
Auto, normal (triggered), single
~60 ns to 10 seconds
15 ps rms
(continued)