53132A/ABA Agilent Technologies Test Equipment, 53132A/ABA Datasheet - Page 4

no-image

53132A/ABA

Manufacturer Part Number
53132A/ABA
Description
COUNTER, 225 MHZ UNIVERSAL COUNTER 12 DIGITS/S
Manufacturer
Agilent Technologies Test Equipment
Datasheet
Instrument Inputs
Input Specifications
Channel 1 & 2 (53131A, 53132A)
Channel 1 (53181A)
Frequency Range
dc Coupled
ac Coupled
FM Tolerance 25%
Voltage Range and Sensitivity (Sinusoid)
dc to 100 MHz 20 mVrms to ±5 V ac + dc
100 MHz
to 200 MHz
200 MHz
to 225 MHz
Voltage Range and Sensitivity
(Single-Shot Pulse)
4.5 ns to 10 ns 100 mVpp to 10 Vpp
Pulse Width
>10 ns
Pulse Width
Trigger Level
Range
Accuracy
Resolution
Damage Level
50 Ω
0 to 3.5 kHz,
1 MΩ
3.5 kHz to
100 kHz, 1 MΩ derated to 5 Vrms
>100 kHz,
1 MΩ
2
dc to 225 MHz
1 MHz to 225 MHz (50 Ω)
30 Hz to 225 MHz (1 MΩ)
30 mVrms to ±5 V ac + dc
40 mVrms to ±5 V ac + dc
(all specified at 75 mVrms
with opt. rear connectors)
(150 mVpp with optional
rear connectors)
50 mVpp to 10 Vpp
(100 mVpp with optional
rear connectors)
± 5.125 V
± (15 mV + 1% of trigger level)
5 mV
5 Vrms
350 Vdc + ac pk
350 Vdc + ac pk linearly
5 Vrms
2
3
3
1
3
2
Input Characteristics
Channel 1 & 2 (53131A, 53132A)
Channel 1 (53181A)
Impedance
1 MΩ
Capacitance
Coupling
Low-Pass Filter 100 kHz, switchable
Input
Sensitivity
Trigger Slope Positive or Negative
Auto Trigger Level
Range
Frequency
Input Amplitude > 100 mVpp
Attenuator
Voltage Range x10
Trigger Range x10
Input Specifications
Channel 3 (53131A, 53132A)
Channel 2 (53181A)
Frequency Range
Option 015
(for 53181A
only)
Option 030
Option 050
Option 124
Power Range and Sensitivity (Sinusoid)
Option 030
Option 050
Option 124
Damage Level
Option 030
Option 050
Option 124
1 MΩ or 50 Ω
30 pF
ac or dc
-20 dB at > 1 MHz
Selectable between Low,
Medium, or High (default).
Low is approximately 2x
High Sensitivity.
0 to 100% in 10% steps
> 100 Hz
(No amplitude modulation)
100 MHz to 1.5 GHz
(see Opt. 030 for
additional specs)
100 MHz to 3 GHz
200 MHz to 5 GHz
200 MHz to 12.4 GHz
100 MHz to 2.7 GHz:
2.7 GHz to 3 GHz:
200 MHz to 5 GHz:
200 MHz to 12.4 GHz
-27 dBm to +19 dBm
-21 dBm to +13 dBm
-23 dBm to +13 dBm
-23 dBm to +13 dBm
5 Vrms
+25 dBm
+25 dBm
4
4
1
Characteristics
Impedance
Coupling
VSWR
External Arm Input Specifications
Signal Input Range
TTL Compatible
Timing Restrictions
Pulse Width
Transition Time
Start-to-Stop Time > 50 ns
Damage Level
External Arm Input Characteristics
Impedance
Input Capacitance 17 pF
Start/Stop Slope Positive or Negative
External Time Base Input Specifications
Voltage Range
Damage Level
Frequency
Time Base Output Specifications
Output Frequency 10 MHz
Voltage
1. Specifications and Characteristics for Channels
2. Values shown are for X1 attenuator setting.
3. When the 53131A or 53132A are ordered with
4. When optional additional channels are ordered
5. Available for all measurements except Peak Volts.
1 and 2 are identical for both common and
separate configurations.
Multiply all values by 10 (nominal) when using
the X10 attenuator setting.
the optional rear terminals (Opt. 060), the channel
1 and 2 inputs are active on both front and rear
of the counter. When the 53181A is ordered with
the optional rear terminal, the channel 1 input is
active on both front and rear of the counter. For
this condition, specifications indicated for the rear
connections also apply to the front connections.
with Opt. 060, refer to configuration table for
Opt. 060 under ordering info on page 8. There is
no degradation in specifications for this input,
as applicable.
External Arm is referred to as External Gate for
some measurements.
50 Ω
AC
<2.5:1
> 50 ns
< 250 ns
10 Vrms
1 kΩ
200 mVrms to 10 Vrms
10 Vrms
1 MHz, 5 MHz, and 10 MHz
(53132A 10 MHz only)
> 1 Vpp into 50 Ω
(centered around 0 V)
5
5