E4416A Agilent Technologies Test Equipment, E4416A Datasheet - Page 3

no-image

E4416A

Manufacturer Part Number
E4416A
Description
Meter, Power
Manufacturer
Agilent Technologies Test Equipment
Type
Wattr
Datasheet
Absolute accuracy (log)
Absolute accuracy (linear) ±0.8%
Relative accuracy (log)
Relative accuracy (linear)
1. Power meter is within ±5 °C of its calibration temperature.
2. National metrology institutes of member states of the Metre Convention, such as the
3. Fast speed is not available for 8480 and N8480 series sensors.
4. Maximum measurement speed is obtained by using binary output in free run trigger.
5. For E9320 sensors, maximum speed is achieved using binary output in free run
Accuracy
Instrumentation
Please add the corresponding power sensor linearity per-
centage; see Tables 6a and 6b for the E9320 sensors.
Average only mode:
Absolute
Relative
Normal mode:
Time Base Accuracy
1 mW power reference
Power output:
Accuracy:
Frequency:
SWR:
Connector type:
National Institute of Standards and Technology in the USA, are signatories to the
ComitÈ International des Poids et Mesures Mutual Recognition Arrangement. Further
information is available from the Bureau International des Poids et Mesures,
acquisition.
For two years
Calibration temperature
±5 °C
±0.04 dB
±0.08 dB
±1.6%
Logarithmic: ±0.02 dB
Linear: ±0.5%
Logarithmic: ±0.04 dB
Linear: ±1.0%
0.01%
1.00 mW (0.0 dBm). Factory
set to ±0.4% traceable to the
National Physical Laboratories
(NPL), UK
±0.5% (23 ± 3 °C)
±0.6% (25 ± 10 °C)
±0.9% (0 to 55 °C)
50 MHz nominal
1.06 maximum (1.08 maximum
for Option E41xA-003)
Type N (f), 50 ohms
2
1
Temperature
0 to 55 °C
±0.08 dB
±1.7%
±0.16 dB
±3.4%
Measurement characteristics:
Measurements:
Averaging:
Measurement speed (GPIB)
Over the GPIB, three measurement speeds are available
(normal, x 2 and fast). The typical maximum speed is
shown in the table below.
Table 2. Measurement speed for different sensor types
Channel functions
Storage registers
Predefined setups
For common wireless standards (GSM900, EDGE, NADC,
iDEN, Bluetooth, IS-95 CDMA, W-CDMA and cdma2000),
predefined setups are provided.
E-Series E9320
peak and average
sensors
E-Series CW and E9300 average power
sensors
8480 and N8480 Series sensor
Sensor type
Average only mode
Normal mode
Average power
Peak power
Peak-to-average ratio
Measurements between two time
offsets (time-gating)
Averaging over 1 to 1024 readings
is available for reducing noise
A, B, A/B, B/A, A-B, B-A and
Relative
10 instrument states can be
saved via the Save/Recall menu.
5
Normal
Measurement speed
20
20
(readings/second)
20
20
x 2
40
40
40
40
Fast
1000
400
400
N.A.
3,4
3