LG01 CONNECTIVITY KIT SENSIRION, LG01 CONNECTIVITY KIT Datasheet - Page 2

no-image

LG01 CONNECTIVITY KIT

Manufacturer Part Number
LG01 CONNECTIVITY KIT
Description
CONNECTIVITY KIT, FOR LG01 SERIES
Manufacturer
SENSIRION
Datasheet

Specifications of LG01 CONNECTIVITY KIT

Accessory Type
Conection Kit
Rohs Compliant
Yes
For Use With
LG01 Sensors
LG01 Liquid Flow Switch
3
3.1 Electrical Specifications
An electrical voltage output indicates flow above (5 V) or below (0 V) the
inside the sensors flow
calibration for other media is available on request-
Table 3: DC Characteristics.
3.2 Electrical Connector and Pinout
Connector type on PCB: 5 pin Molex PCB Header 53261-0590. Mating connector: Molex 51021-0500.
3.3 Mechanical Specifications and Pressure Rating
Inside the LG01 flow sensors a highly sensitive microsensor-chip is mounted on the outside of a straight glass
capillary and allows to measure the flow through the wall (US Patent 6,813,944 B2) using a thermal principle.
Table 4: Mechanical Specifications and Pressure Rating
a
to Barb connectors are available from company Value Plastics as well.
www.sensirion.com
For fluid connection material use flat bottom ¼-28 standard components e.g. IDEX parts P-200/P-300. Various adapters e.g.
Parameter
Power Supply DC, VDD
Operating Current
Voltage Output High
Voltage Output Low
Attention Fragile
Parameter
Fluid Connector Ports (Fittings)
Wetted Materials:
Overpressure Resistance
Maximum Pressure Drop (at max. flow rate)
Internal Sensor Capillary, Inner Diameter
Total Internal Volume
Total Mass
suitable tools causes stress on the internal thin-walled glass capillary.
This can cause it to break! Test for leakage every time after new connections are made.
Electrical and Mechanical Specifications
Forces introduced by stiff fluid connection lines as well as connecting the fittings without
Pin
1
2
3
4
5
Internal Sensor Capillary Material
Fitting Material
Additional Sealing Material
do not connect
do not connect
VDD
GND
Voltage Output
channel. The product comes fully calibrated for water – for volume applications flow
Flow>Flow Switch Limit and no air
Flow<Flow Switch Limit or air found
Conditions
V
DD
= 9 V, no load
1
2
3
May_ 2011 _ V1.3
4 5
for 1/16” or 1/8” OD plastic tubing
100% PEEK™ (polyetheretherketone)
Borosilicate Glass 3.3
ETFE(Tefzel
Flow Switch Limit and/or air or bubbles
LG01-2000
Min.
(Duran
1.8 mm
1 mbar
<90 µl
¼-28
40 psi
5
3 bar
6 g
®
)
®
)
Typ.
6.8
7
5
0
Max.
12
a
a
Units
mA
V
V
V
!
2/4