FLUKE-8846A NIST W/O DATA Fluke, FLUKE-8846A NIST W/O DATA Datasheet - Page 3

no-image

FLUKE-8846A NIST W/O DATA

Manufacturer Part Number
FLUKE-8846A NIST W/O DATA
Description
Digital Multimeters 6.5 BENCH DMM NIST CAL CERT
Manufacturer
Fluke

Specifications of FLUKE-8846A NIST W/O DATA

Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
1259800
2
Electrical
Input Protection ............................................................1000 V all ranges
Overrange.......................................................................20 % on all ranges except 1000 V dc, 1000 V ac (8846A),
Remote Interfaces
Warranty
Accuracy specifications are valid for 6½ digit resolution mode after at least a 1-hour warm-up with Auto Zero enabled.
24-hour specifications are relative to calibration standards and assume a controlled electromagnetic environment per EN 61326-1:2000-11
DC Voltage Specifications
Maximum Input.............................................................1000 V on any range
Common Mode Rejection............................................140 dB at 50 or 60 Hz ±0.1 % (1 kΩ unbalance)
Normal Mode Rejection ..............................................60 dB for NPLC of 1 or greater with analog filter off and power line frequency ±0.1 %
Measurement Method..................................................Multi-ramp A/D
A/D Linearity .................................................................0.0002 % of measurement +0.0001 % of range
Input Bias Current ........................................................<30 pA at 25 °C
Autozero Off Operation ...............................................Following instrument warm-up at calibration temperature ±1 °C and less than 10
Analog Filter..................................................................When using the analog filter, specifications are relative to within one hour of using the
DC Ratio ..........................................................................Accuracy is +/- (Input accuracy + Reference accuracy), where Input accuracy = DC
Settling Considerations ..............................................Measurement settling times are affected by source impedance, cable dielectric
Input Characteristics
8846A Accuracy
Accuracy is given as ± (% measurement + % of range)
Fluke Corporation
100 mV
1 V
10 V
100 V
1000 V
100 mV
1 V
10 V
100 V
1000 V
Range
[1]
RS-232C, DTE 9-pin, 1200 to 230400 baud (RS-232C to USB cable available to connect the Meter to a PC USB port. See Accessories)
IEEE 488.2
LAN and “Ethernet 10/100 base T with DHCP (for IP_ADDRess) option”
One year
Range
Inputs beyond ±14 V are clamped through 200 kΩ typical. 10 MΩ is default input impedance.
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1,000.000 V
Resolution
0.0025 + 0.003
0.0018 + 0.0006
0.0013 + 0.0004
0.0018 + 0.0006
0.0018 + 0.0006
8845A and 8846A
(23 ±1 °C)
24 Hour
10 mV
100 mV
10 µV
100 µV
1 mV
4½ Digits
0.0025 + 0.0035
0.0018 + 0.0007
0.0018 + 0.0005
0.0027 + 0.0006
0.0031 + 0.001
(23 ±5 °C)
90 Days
750 V ac (8845A), Diode, and 10 A ranges
100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1 %
minutes, add error: 0.0002 % range additional error + 5 μV.
ZERO function for that range and NPLC setting.
Voltage accuracy for the HI to LO Input (in ppm of the Input voltage), and Reference
accuracy = DC Voltage accuracy for the HI to LO (Sense) Reference (in ppm of the
Reference voltage).
characteristics, and input signal changes.
1 µV
10 µV
100 µV
1 mV
10 mV
Resolution
5½ Digits
0.0037 + 0.0035
0.0025 + 0.0007
0.0024 + 0.0005
0.0038 + 0.0006
0.0041 + 0.001
(23 ±5 °C)
1 Year
Electrical Specifications
100 nV
1 µV
10 µV
100 µV
1 mV
Outside 18 to 28 °C
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
6½ Digits
Coefficient/ °C
Temperature
10 MΩ or >10 GΩ
10 MΩ or >10 GΩ
10 MΩ or >10 GΩ
10 MΩ ±1%
10 MΩ ±1%
Input Impedance
[1]