433202-NIST EXTECH INSTRUMENTS, 433202-NIST Datasheet

Clamp Multimeters & Accessories CALIBRATOR WITH NI

433202-NIST

Manufacturer Part Number
433202-NIST
Description
Clamp Multimeters & Accessories CALIBRATOR WITH NI
Manufacturer
EXTECH INSTRUMENTS
Datasheet

Specifications of 433202-NIST

Accuracy
+/- 0.15 %
Ranging
9 V
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
EXTECH INSTRUMENTS CORPORATION
285 Bear Hill Road, Waltham, MA 02451-1064, U.S.A.
Phone: 781-890-7440 • FAX 781-890-7864
ISO 9001:2000 CERTIFIED
Microprocessor Calibrator Thermometer
Source and Measure 8 thermocouple type devices J, K, T, E, C, R, S and N
Specifications:
Type J
Type K
Type T
Type E
Type C
Type R
Type S
Type N
Voltage
Resolution:
Cold Junction Compensation:
Input Impedance:
Sampling Time:
Dims/Weight:
Features:
One model for multiple thermocouple types with
precision output displayed as mV or °C/°F
Basic accuracy of 0.15% of reading over wide ranges
Five stored calibration values for fast calibrations
High accuracy calibration function simulates precision
thermocouple outputs for use in calibrating thermometers,
transmitters, controllers or recorders.
0.1°/1° resolution for J, K, T and E Types; 1° for C, R, S and N Types
°C/°F switchable for increased versatility
Displays output in terms of millivolts or temperature
based on thermocouple tables
4 digit LCD display is easy to read with thermocouple type indication
Unique handheld Oyster case design features large digital display
built into “flip-up" cover automatically shuts off when closed
Portable and rugged industrial design for field, plant or benchtop use
Operates on alkaline batteries, rechargeable batteries or AC adaptor
Standard Thermometer Calibration cables (terminated with
subminiature connectors on either side) comes standard
with the meter. Optional Process Calibration Cable is
terminated with spade lug for process applications
Complete with Type K Thermocouple calibration cable, Copper calibration
cable, 9V NiCad rechargeable battery, neckstrap for "hands-free"
operation, AC adaptor/recharger, and hard carrying case
Ranges for Calibrate and Measure modes
-58 to 1830°F
-58 to 2498°F
-184 to 752°F
-58 to 1382°F
32 to 4172°F
32 to 3182°F
32 to 3182°F
-58 to 2372°F
-5.00mV to +55.00mV
3.8x4.2x1.8" (96x108x45mm)/ 17oz
(-50 to 1000°C)
(-50 to 1370°C)
(-270 to 400°C)
(-50 to 750°C)
(0 to 2300°C)
(0 to 1750°C)
(0 to 1750°C)
(-50 to 1300°C)
10µV
0.1° (up to 999.9) or 1° (over 999.9°)
0.03°C/°C (0.02°F/°F)
10 Mohm
4 times/second
Product
Accuracy
0.15% rdg ±1°
0.15% rdg ±1°
0.15% rdg ±1°
0.15% rdg ±1°
0.15% rdg ±1°
0.15% rdg ±1°
0.15% rdg ±1°
0.15% rdg ±1°
10µV ±1d
Copyright © 2005 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form.
Ordering Information:
433201 ............Microprocessor Calibrator Thermometer (115V)
433201-NIST ..Calibrator Thermometer (115V) w/NIST Certificate
433202 ............Microprocessor Calibrator Thermometer (220V)
433202-NIST ..Calibrator Thermometer (220V) Certificate
431437 ............Optional Process Calibration Cable with Spade Lugs
409996 ............Vinyl Pouch Carrying Case
“Make mine an Extech!
www.extech.com
DATASHEET
05/04/06 - R1