FLUKE-1587-ET Fluke, FLUKE-1587-ET Datasheet - Page 5

Digital Multimeters ADVANCED ELECTRICAL TROUBLESHOOTING KIT

FLUKE-1587-ET

Manufacturer Part Number
FLUKE-1587-ET
Description
Digital Multimeters ADVANCED ELECTRICAL TROUBLESHOOTING KIT
Manufacturer
Fluke
Datasheet

Specifications of FLUKE-1587-ET

Accuracy
600 mV to 1000 V
Resistance Range
50 M Ohms
Capacitance Range
1000 nF, 10 uF, 100 uF, 9999 uF
Equipment Type
Combo Kits
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
2509646
1587/1577 General Specifications
Model 1587
Model 1577
Maximum voltage applied to any terminal: 1000 V ac rms
or dc
Storage temperature: -40 °C to 60 °C (-40 °F to 140 °F)
Operating temperature: -20 °C to 55 °C (-4 °F to 131 °F)
Temperature coefficient: 0.05 x (specified accuracy) per °C
for temperatures < 18 °C or > 28 °C (< 64 °F or > 82 °F)
Relative humidity, non-condensing: < °C
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
Vibration: Random, 2 g, 5-500 Hz per MIL-PRF-28800F,
Class 2 instrument
Shock: 1 meter drop per IEC 61010-1 2
drop test, six sides, oak floor)
Electromagnetic compatibility: In an RF field of 3 V/M,
accuracy = specified accuracy except in temperature:
accuracy = specified accuracy ± 5 °C (9 °F). (EN 61326-1:1997)
Safety: Complies with ANSI/ISA 82.02.01 (61010-1) 2004,
CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2
Edition for measurement CAT III 1000 V and CAT IV 600 V
(0 % to + 20 %)
(0 % to + 20 %)
(0 % to + 20 %)
(0 % to + 20 %)
(0 % to + 20 %)
(0 % to + 20 %)
(0 % to + 20 %)
Output Voltage
Output Voltage
1000 V
1000 V
250 V
500 V
500 V
100 V
50 V
0.01 to 6.00 MΩ
0.01 to 6.00 MΩ
Display Range
Display Range
6.0 to 50.0 MΩ
6.0 to 60.0 MΩ
0.1 to 60.0 MΩ
0.1 to 60.0 MΩ
0.1 to 60.0 MΩ
0.1 to 60.0 MΩ
0.1 to 60.0 MΩ
60 to 100 MΩ
60 to 250 MΩ
60 to 500 MΩ
60 to 600 MΩ
60 to 500 MΩ
60 to 600 MΩ
0.6 to 2.0 GΩ
nd
Edition (1 meter
Resolution
Resolution
0.01 MΩ
0.01 MΩ
100 MΩ
0.1 MΩ
0.1 MΩ
0.1 MΩ
0.1 MΩ
0.1 MΩ
0.1 MΩ
0.1 MΩ
nd
1 MΩ
1 MΩ
1 MΩ
1 MΩ
1 MΩ
1 MΩ
Certifications: CSA per standard CSA/CAN C22.2
No. 61010.1-04; TUV per standard EN 61010 Part 1-1002
Batteries: Four AA batteries (NEDA 15A or IEC LR6)
Battery life: Meter use 1000 hours; Insulation test use: Meter
can perform at least 1000 insulation tests with fresh alkaline
batteries at room temperature. These are standard tests of
1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25
seconds off
Size: 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in
W x 8.00 in L)
Weight: 550 g (1.2 lb)
IP rating: IP40
Altitude (operating): 2000 m CAT III 1000 V, CAT IV 600 V;
3000 m CAT II 1000 V, CAT III 600 V
Storage: 12,000 m
Over-range capability: 110 % of range except for
capacitance which is 1 %
Compliance to EN 61557: IEC61557-1, IEC61557-2
1 mA @ 250 kΩ
1 mA @ 500 kΩ
1 mA @ 500 kΩ
1 mA @ 100 kΩ
1 mA @ 50 kΩ
1 mA @ 1 MΩ
1 mA @ 1 MΩ
Test Current
Test Current
1587/1577 Insulation Multimeters
Resistance Accuracy
Resistance Accuracy
± (2.0 % + 5 counts)
± (2.0 % + 5 counts)
± (1.5 % + 5 counts)
± (1.5 % + 5 counts)
± (1.5 % + 5 counts)
± (10 % + 3 counts)
± (3 % + 5 counts)
± (3 % + 5 counts)
Fluke Corporation 5