DSO5052A AGILENT TECHNOLOGIES, DSO5052A Datasheet - Page 16

OSCILLOSCOPE, 500MHZ, 2 CHANNEL, 4GSPS

DSO5052A

Manufacturer Part Number
DSO5052A
Description
OSCILLOSCOPE, 500MHZ, 2 CHANNEL, 4GSPS
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of DSO5052A

Scope Type
Digital
Scope Channels
2 Analog
Bandwidth
500MHz
Meter Display Type
TFT-LCD Color
Sampling Rate
4GSPS
Input Impedance
1Mohm
Rise Time
700ps
Input Voltage
400V P-p
Rohs Compliant
NA
16
Performance characteristics
Trigger system (continued)
Selections
AutoScale
Edge
Pulse width
Pattern
Sequence
TV
Duration
CAN
FlexRay
LIN
I
SPI
RS-232/UART
MIL-STD 1553
2
C
Edge, pulse width, pattern, TV, duration
Trigger on a rising, falling, or alternating edge of any source
Trigger when a positive- or negative-going pulse is less than, greater than, or within a specified range
on any of the source channels.
Minimum pulse width setting:
Maximum pulse width setting: 10 s
Trigger at the beginning of a pattern of high, low, and don't care levels and/or a rising or falling edge
established across any of the channels, but only after a pattern has been established for a minimum of
2 nsec. The channel’s high or low level is defined by that channel’s trigger level.
Arm on Event A, trigger on Event B(edge or pattern), with option to reset on Event C or time delay
Trigger using any oscilloscope channel on most analog progressive and interlaced video standards
including HDTV/EDTV, NTSC, PAL, PAL-M or SECAM broadcast standards. Select either positive or
negative sync pulse polarity. Modes supported include Field 1, Field 2, all fields, all lines, or any line
within a field. TV trigger sensitivity: 0.5 division of sync signal. Trigger holdoff time can be adjusted in
half field increments.
Trigger on a multi-channel pattern whose time duration is less than a value, greater than a value,
greater than a time value with a timeout, or inside or outside of a set of time values.
Minimum duration setting: 2 ns
Maximum duration setting: 10 s
Trigger on CAN (Controller Area Network) version 2.0A and 2.0B signals. Trigger on the start of frame
(SOF) bit (standard). N5424A option supports triggering on remote frame ID (RTR), data frame ID
(~RTR), remote or data frame ID, data frame ID and data, error frame, all errors, acknowledge error and
overload frame.
Trigger on FlexRay Frames, errors, events and cycle-multiplexed triggering. N5432C or option
FLX supports also triggering on particular frame types symbolically, such as Startup frames, Null frame,
Sync frame, etc., as well as Boolean NOT frame types.
Trigger on LIN (Local Interconnect Network) sync break at beginning of message frame (standard).
N5424A option supports triggering on frame ID.
Trigger on I
and/or data values. Also trigger on missing acknowledge, address with no acq, restart, EEPROM read,
and 10-bit write.
Trigger on SPI (Serial Protocol Interface) data pattern during a specific framing period. Supports positive
and negative Chip Select framing as well as clock Idle framing and user-specified number of bits per
frame.
View the information sent over a RS-232 serial bus. Display real-time time-aligned decode of transmit
and receive lines. The application also enables triggering on RS-232/UART conditions.
Trigger on specific Command/Status Words, Data Words, and error conditions.
Finds and displays all active channels, sets edge trigger mode on highest-numbered channel, sets
vertical sensitivity on channels, time base to display ~1.8 periods. Requires minimum voltage >
10 mVpp, 0.5% duty cycle and minimum frequency > 50 Hz.
5 ns (DSO501xA)
2 ns (DSO503xA, DSO505xA)
2
C (Inter-IC bus) serial protocol at a start/stop condition or user defined frame with address
(continued)