Datasheet

Specifications of DSOX2014A-GSA

Scope Type
Digital Bench
Scope Channels
4 Analog
Bandwidth
100MHz
Meter Display Type
WVGA Color
Sampling Rate
2GSPS
Input Impedance
1Mohm
Input Voltage
300V
External Height
8.05"
Rohs Compliant
NA
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characterisitics
* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and from ±10 °C firmware calibration temperature
Waveform math
Operators
FFT
Sources
Display characteristics
Display
Resolution
Interpolation
Persistence
Intensity gradation
Modes
MSO (digital channels)
Upgradable from DSO
MSO channels
Maximum sample rate
Maximum record length
Threshold selections
Threshold accuracy*
Maximum input dynamic
range
Minimum voltage swing
Input impedance
Minimum detectable pulse
width
Channel-to-channel skew
Environmental and safety
Power line consumption
Temperature
Humidity
Altitude
Electromagnetic
compatibility
Safety
Add, subtract, multiply, FFT
Windows: Hanning, flat top, rectangular; Blackman-Harris - up to 64 kpts resolution
Math functions available between any two channels
8.5-inch WVGA
800 (H) x 480 (V) pixel format (screen area)
Sin(x)/x interpolation (using FIR filter; used when there is less than one sample per column of the display)
Off, infinite, variable persistence (100ms-60s)
64 intensity levels
Normal
XY - XY mode changes the display from voltage versus time scale to a volts versus volts scale
Roll – Displays the waveform moving across the screen from right to left much like a strip chart recorder
Yes
8 channels (D0 to D7)
1 GSa/s
50kpts per channel (digital channels only)
12.5kpts per channel (analog and digital channels)
TTL (+1.4 V), CMOS (+2.5 V), ECL (-1.3 V), User-definable (±8.0 V in 10 mV stops)
± (100 mV + 3% of threshold settings)
±10 V about threshold
500 mVpp
100 kΩ ± 2% at probe tip, ~8 pF
5ns
2 ns (typical), 3 ns (maximum)
100 watts
Operating: 0 to +55 °C
Non-operating: -30 to +71 °C
Operating: Up to 80% RH at or below +40 °C; up to 45% RH up to +50 °C
Non-operating: Up to 95% RH up to 40 °C; up to 45% RH up to 50 °C
Operating and non-operating: up to 4,000 m
Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS
UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04