ZS-24-A SMC Corporation of America, ZS-24-A Datasheet
ZS-24-A
Specifications of ZS-24-A
Related parts for ZS-24-A
ZS-24-A Summary of contents
Page 1
... M5 x 0.8 Input/Output specifications Setting range Lead wire length –100.0 to 100.0 kPa –101.3 to 101.3 kPa –1.000 to 1.000 MPa ZSE40/ISE40 Option Nil None Bracket A (ZS-24-A) A Bracket B (ZS-24-B) B Panel mount (ZS-22-A) E Panel mount (ZS-24-C) + Front protective cover F ∗ When optional parts only are required, order with the part numbers inside ( ) ...
Page 2
... Note 2) For ZSE40 (F)/ISE40- Note 3) For ZSE40 (F)/ISE40-- 70 Note 4) For ZSE40F (compound pressure) with “psi” indication, this is 0.03 to 0.04 psi. Note 5) For ZSE40F (compound pressure) with “psi” indication, zero clear is in the range of ±0.01 psi. Example of Internal Circuit and Wiring ZSE40(F) ISE40- ...
Page 3
... Switches to each mode and fixes the set value. Key lock mode Mode is not switched, even if the button is pressed during operation. Normal operation Measured pressure displayed and the switch operation begins. ZSE ISE PSE Z SE3 ZSP ISA2 IS ZSM PF2 IF Data is 16-2-19 ...
Page 4
... Display of setting auto shift SET “SET” button. Press the “SET” button to complete the setting. ∗ Only for ZSE40(F)/ISE40--30/70(L)-M, displays the input mode “ ” and the compensation value alternately. If auto shift input is not complete, zero is displayed. ...
Page 5
... Unit Set It's only subject to ZSE40(F)/ISE40--(L). Unit selection Set the unit by either button or button. : kPa or MPa, : kgf/ bar, : psi, : inHg, : mmHg SET Press “SET” button. OUT1: Selection of the output mode Auto Preset (For adsorption confirmation) 1. Auto preset mode ...
Page 6
... ZSE40/ISE40 Series Calibration Procedures Auto Preset (In the case of confirming the supply pressure) 1. Auto preset mode Select the Auto preset mode in the initial setting mode and press the “SET” button until “ ” is displayed. ON point (C, C1) Hysteresis (3 digits) OFF ON point = C, C1 OFF point = C-3 digits, C1-3 digits (1 digit is the minimum setting unit ...
Page 7
... For splash proof use (IP65), insert an air tube into the atmospheric release port. (Refer to “Precautions” on page 16-2-24 for details.) ZSE40/ISE40 20 ZSE ISE PSE Z SE3 ZSP ISA2 IS ZSM PF2 IF Data 20 Rc 1/8 16-2- ...
Page 8
... ZSE40/ISE40 Series Dimensions C4 ZSE40 ( F ) /ISE40– 32 MPa OUT1 OUT2 SET Atmospheric release port 7.8 12 One-touch fitting ø4, ø6 ∗ For splash proof use (IP65), insert an air tube into the atmospheric release port. (Refer to “Precautions” for details.) Caution 1. Immediately after supplying power, there is drift of about ±0.5% F.S. When used with very low pressure, allow the unit to warm up for about minutes ...