2306-US KEITHLEY, 2306-US Datasheet - Page 7

58T8426

2306-US

Manufacturer Part Number
2306-US
Description
58T8426
Manufacturer
KEITHLEY
Datasheet

Specifications of 2306-US

No. Of Outputs
2
Output Voltage
15V
Output Voltage 2
15V
Output Current
5A
Output Current 2
5A
Load Regulation
0.01%
Length
411mm
Width
213mm
Height
89mm
Weight
3.9kg
Power Supply Output Type
Variable
Input Voltage
100VAC To 120VAC / 220VAC To 240VAC
Power Rating
50W
2306-VS Dual Channel Battery/Charger Simulator
10 Typical values, peak-to-peak noise equals 6 times rms noise.
11 Based on settled signal: 100µs pulse trigger delay.
12 Also applies to other apertures that are integer multiples of 1PLC.
13 Recovery to within 20mV of previous level.
Specifications are subject to change without notice.
1 PLC = 1.00.
2 Following 15 minute warm-up, the change in output over 8 hours under ambient
3 Remote sense, at output terminals, 0.5A to 5A typical.
4 Remote sense, with 4.5m (15 ft) of 16 gauge (1.31mm2) wire and 1Ω resistance in
5 Minimum current in constant current mode is 6mA.
6 60Hz (50Hz).
7 PLC = Power Line Cycle. 1PLC = 16.7ms for 60Hz operation, 20ms for 50Hz
8 Display off.
9 Speed includes measurement and binary data transfer out of GPIB.
ISOLATION (LOW–EARTH): 22VDC max. Do not exceed 60VDC between any
PROGRAMMING: IEEE-488.2 (SCPI).
USER-DEFINABLE POWER-UP STATES: 3.
REAR PANEL CONNECTORS: Two trigger in and two trigger out (BNC)
TRIGGER IN/OUT CONNECTORS: IN High 3-5V, IN Low ≤0.8V, OUT High
TEMPERATURE COEFFICIENT (OUTSIDE 23°C ±5°C): Derate accuracy
OPERATING TEMPERATURE: 0° to 50°C (Derate to 70%). 0° to 35°C (Full
STORAGE TEMPERATURE: –20° to 70°C.
HUMIDITY: <80% @ 35°C non-condensing.
DISPLAY TYPE: 2-line x 16 character VFD.
DIMENSIONS: 89mm high x 213mm wide x 411mm deep (31¼2 in x 83¼8 in x
NET WEIGHT: 3.9kg (8.6lbs.)
SHIPPING WEIGHT: 6.4kg (14lbs.)
INPUT POWER: 100–120VAC/220–240VAC, 50 or 60Hz (auto detected at power-
POWER CONSUMPTION: 165VA max.
WARRANTY: Two years parts and labor on materials and workmanship.
EMC: Conforms with European Union Directive directive 89/336/EEC, EN 61326.
SAFETY: Conforms with European Union Directive 73/23/EEC, EN 61010-1.
VIBRATION: MIL-PRF-28800F Type III, Class 3
ACCESSORIES SUPPLIED: User and service manual, output connectors mating
temperature, constant load, and line operating conditions.
each lead to simulate typical test environment, 1.5A load change (0.15A to 1.65A).
operation.
two terminals of either connector.
connectors. Two 8-position quick disconnect terminal block for output (4), sense
(2), and DVM (2).
>4V, OUT Low <0.8V.
specification by (0.1 x specification)/°C.
power).
163¼16 in).
up).
terminal (part no. CS-846).
GENERAL
HW 6/20/03
Page 7 of 7
Rev. B

Related parts for 2306-US