MDO4054-3GSA Tektronix, MDO4054-3GSA Datasheet - Page 13

no-image

MDO4054-3GSA

Manufacturer Part Number
MDO4054-3GSA
Description
Specifications: Scope Type: Bench ; Scope Channels: 4 Analogue + 16 Digital + 1 RF ; Bandwidth: 500MHz ; Meter Display Type: XGA TFT-LCD Color ; Sampling Rate: 2.5GSPS ; Input Impedance: 1Mohm ; Rise Time: 700ps ; Input Voltage: 300Vrms ; Supply Volt
Manufacturer
Tektronix
Datasheet
Trigger System
Characteristic
Main Trigger Modes
Trigger Coupling
Trigger Holdoff
Range
Trigger Modes
Mode
Edge
Sequence (B-trigger)
Pulse Width
Timeout
Runt
Logic
Setup and Hold
Rise/Fall Time
Video
Extended Video
(Optional)
I
SPI (Optional)
2
C (Optional)
Trigger on violations of both setup time and hold time between clock and data present on one or more of any of the analog and digital input channels
Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative, or either
Description
Positive or negative slope on any channel. Coupling includes DC, AC, HF reject, LF reject, and noise reject
Trigger Delay by Time – 4 ns to 8 s. Or Trigger Delay by Events – 1 to 4,000,000 events
Trigger on width of positive or negative pulses that are >, <, =, or ≠ a specified period of time (4 ns to 8 s)
Trigger when no pulse is detected within a specified time (4 ns to 8 s)
Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again
Trigger when any logical pattern of channels goes false or stays true for specified period of time (4 ns to 8 s). Any input can be used as a clock to
look for the pattern on a clock edge. Pattern (AND, OR, NAND, NOR) specified for all analog and digital input channels defined as High, Low, or
Don't Care
Trigger on all lines, odd, even, or all fields on NTSC, PAL, and SECAM video signals
Trigger on 480p/60, 576p/50, 720p/30, 720p/50, 720p/60, 875i/60, 1080i/50, 1080i/60, 1080p/24, 1080p/24sF, 1080p/25, 1080p/30, 1080p/50,
1080p/60, and custom bi-level and tri-level sync video standards
Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data, or Address and Data on I
Trigger on SS active, MOSI, MISO, or MOSI and MISO on SPI buses up to 50 Mb/s
Description
Auto, Normal, and Single
DC, AC, HF reject (attenuates >50 kHz), LF
reject (attenuates <50 kHz), noise reject (reduces
sensitivity)
20 ns to 8 s
Trigger Sensitivity
Characteristic
Internal DC Coupled
1 MΩ Path
(All models)
50 Ω Path
(MDO4054-X)
50 Ω Path
(MDO4104-X)
Trigger Level Range
Characteristic
Any Channel
Line
Trigger Frequency Readout
Provides 6-digit frequency readout of triggerable events.
Mixed Domain Oscilloscopes — MDO4000 Series
Description
For 1 mV/div to 4.98 mV/div; 0.75 div from DC to
50 MHz, increasing to 1.3 div at rated bandwidth
For ≥5 mV/div; 0.4 div from DC to 50 MHz,
increasing to 1 div at rated bandwidth
0.4 div from DC to 50 MHz, increasing to 1 div at
rated bandwidth
Description
±8 divisions from center of screen
Fixed at about 50% of line voltage
2
C buses up to 10 Mb/s
www.tektronix.com
13