FEATURES
This lamp housing was designed to allow
easy operation of deuterium lamps such as
L2D2
performance. It accommodates a lamp with
a flange so that no optical alignment is
required. The built-in interlock and forced-air
cooling functions ensure high safety.
Collimating lenses and fiber guide adaptors
are also available as easy-to-replace
options, which easily attach to the light exit
and allow obtaining the desired light beam.
I Compact
I High Stability
I High Safety, Interlock Function
APPLICATIONS
I Light sources in various experiments
I Light sources in spectrophotometry
SPECIFICATION
NOTE: A A8087 cable should be necessary for connection between C7860 series and E8039.
Optional Parts SMA Fiber Guide Adaptor with Lens (for thin diameter fiber)
Power Supply C7860 series, Connection cable A8087 and Lamp L6999-51 are optional.
Lamp
Output Window
Cooling Function
Operating Ambient Temperature
Recommended Operating Temperature
Operating Humidity
Dimensions ( W × H × D )
Weight
Suitable Power Supply Type No.
DEUTERIUM LAMP HOUSING E8039
Compact
installation.
Lamp housing design allows the L2D2 lamp to fully
achieve its highly stable operation (fluctuation(p-p)
within 0.05%).
The interlock function turns off the lamp if the
cover comes off. The surface temperature of the
lamp housing is maintained below +50 °C.
lamps
Collimating Lens (Material: Fused Silica, Dimension:
case
Parameter
Suitable Lamp
Lamp Power Consumption
Spectral Distribution
Anode Current
Tube Drop Voltage
Material
Dimension
Optical Axis Height
and
allows
provide
easy
handling
full
lamp
and
10 mm, Focal Point: 20 mm, Light output: collimating)
120 × 107 × 140 (exclude projection part)
Forced Air Cooling (Built-in cooling fan)
I No Optical Alignment Required
I Easy Replacement of Optional Parts
Designed to accommodate an L2D2 lamp with a
flange which eliminates arc point shift. Absolutely
no optical alignment is required over the life of the
lamp or when replacing the lamp.
As easy-to-replace options, collimating lenses and
fiber guide adaptors that easily attach to the light
exit are available, allowing you to obtain the
desired light beam.
72 (include feet height)
Description/Value
L2D2 L6999-51
C7860-2510
Fused Silica
Approx. 2.3
Approx. 30
185 to 400
+5 to +35
0 to +40
30 to 90
300
80
10
A