e5gn-rtc Omron Corporation, e5gn-rtc Datasheet - Page 4

no-image

e5gn-rtc

Manufacturer Part Number
e5gn-rtc
Description
Digital Temperature Controllers
Manufacturer
Omron Corporation
Datasheet
■ Characteristics
Note 1. The indication of K thermocouples in the −200 to 1300°C range, and T and N thermocouples at a temperature of −100°C or less, and U
Indication accuracy
Influence of temperature (See note 3.)
Influence of voltage (See note 3.)
Hysteresis
Proportional band (P)
Integral time (I)
Derivative time (D)
Control period
Manual reset value
Alarm setting range
Sampling period
Influence of signal source resistance
Insulation resistance
Dielectric strength
Vibration resistance
Shock resistance
Weight
Degree of protection
Memory protection
EMC
Approved standards
2. “EU” stands for Engineering Unit and is used as the unit after scaling. For a temperature sensor, the EU is °C or °F.
3. Conditions: Ambient temperature: −10°C to 23°C to 55°C, Voltage range: −15% to +10% of rated voltage
and L thermocouples at any temperature is ±2°C±1 digit maximum. The indication of B thermocouples at a temperature of 400°C or less
is unrestricted.
The indication of R and S thermocouples at a temperature of 200°C or less is ±3°C±1 digit maximum.
Thermocouple:
(±0.5% of indicated value or ±1°C, whichever greater) ±1 digit max. (See note 1.)
Platinum resistance thermometer:
(±0.5% of indicated value or ±1°C, whichever greater) ±1 digit max.
Analog input: ±0.5% FS±1 digit max.
CT input: ±5% FS±1 digit max.
R, S, and B thermocouple inputs:
(±1% PV or ±10°C, whichever is greater) ±1 digit max.
Other thermocouple inputs:
(±1% PV or ±4°C, whichever is greater) ±1 digit max.
* K thermocouple at −100°C max.: ±10°C max.
Platinum resistance thermometer inputs:
(±1% PV or ±2°C, whichever is greater) ±1 digit max.
Analog inputs:
(±1% FS) ±1 digit max.
0.1 to 999.9 EU (in units of 0.1 EU) (See note 2.)
0.1 to 999.9 EU (in units of 0.1 EU) (See note 2.)
0 to 3999 s (in units of 1 s)
0 to 3999 s (in units of 1 s)
1 to 99 s (in units of 1 s)
0.0% to 100.0% (in units of 0.1%)
−1999 to 9999 (decimal point position depends on input type)
500 ms
Thermocouple: 0.1°C/Ω max. (100 Ω max.)
Platinum resistance thermometer: 0.4°C(0.8°F)/Ω max. (10 Ω max.)
20 MΩ min. (at 500 VDC)
2000 VAC, 50 or 60 Hz for 1 min (between different charging terminals)
Malfunction: 10 to 55 Hz, 20 m/s
Destruction: 10 to 55 Hz, 0.75-mm amplitude for 2 hours each in X, Y and Z directions
Malfunction: 100 m/s
Destruction: 300 m/s
Approx. 90 g
Front panel: NEMA4X for indoor use (equivalent to IP66), rear case: IP20, terminals: IP00
EEPROM (non-volatile memory) (number of writes: 100,000)
ESD Immunity: EN61000-4-2:
Electromagnetic Immunity: EN61000-4-3:
Burst Noise Immunity: EN61000-4-4:
Surge Immunity: EN61000-4-5:
Conducted Disturbance Immunity: EN61000-4-6:
Voltage Dip/Interrupting Immunity: EN61000-4-11:
Radiated Interference Electromagnetic Field Strength: EN61326 Class A
Noise Terminal Voltage:
UL3121-1, CSA22.2 No. 142, E.B.1402C
Conforms to EN50081-2, EN50082-2, EN61010-1 (IEC61010-1)
Conforms to VDE0106/part 100 (Finger Protection), when the terminal cover is mounted.
2
2
, 3 times each in X, Y, and Z directions
, 3 times each in X, Y, and Z directions
2
for 10 min each in X, Y and Z directions
Mounting bracket: approx. 10 g
4 kV contact discharge
8 kV air discharge
10 V/m (amplitude-modulated, 80 MHz to
1 GHz)
2 kV power line
1 kV I/O signal line
1 kV normal mode (power line, output line
(relay output))
2 kV common mode (power line, output
line (relay output))
3 V (0.15 to 80 MHz)
0.5 cycle, 100% (rated voltage)
EN61326 Class A
E5GN