Setting the Temperature
Maintenance
Calibration and Repair Services
1. Use the up and down arrow buttons to change the set-point value (shown on the
2. Press "SET" and the digits will stop flashing. The calibrator will automatically adjust the
3. When the set-point temperature is changed the controller switches the well heater on
4. Reduce the temperature to a value below 212°F (100°C) before turning the instrument
Warning: Turning off the power at higher temperatures can damage the instrument.
The temperature set-point can be set to any value within the instrument’s specified range.
Do not, for any reason, exceed the specified temperature range.
With proper care the instrument should require very little maintenance.
If the outside of the instrument becomes soiled, it may be wiped clean with a damp cloth
and mild detergent. Do not use harsh chemicals or compounds.
Do not clean the surface of the well with liquids. Simply wipe with a dry cloth.
The calibrator should be handled with care. Avoid dropping or bumping the calibrator.
Extech offers repair and calibration services for the products we sell. Extech also
provides NIST certification for most products. Call the Customer Care Department for
information on calibration services available for this product. Extech recommends that
annual calibrations be performed to verify meter performance and accuracy.
bottom, green LED digit display). The display will flash at a fast rate when the set-point
is changed.
target assembly’s temperature to match the set-point temperature. The red LED
display will indicate the temperature of the target assembly.
or off to raise or lower the temperature. The displayed well temperature gradually
changes until it reaches the set-point temperature. The well may require 5 to 10
minutes to reach the
the stated limits of the set-point. Ultimate stability may take an additional 15 to 20
minutes.
off.
All rights reserved including the right of reproduction in whole or in part in any form.
Product specifications subject to change without notice
Copyright © 2008 Extech Instruments Corporation (a FLIR company)
Technical support: Extension 200; E-mail: support@extech.com
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
For the latest version of this User’s Guide, Software updates, and other
Repair & Returns: Extension 210; E-mail: repair@extech.com
set-point. Another 5 to 10 minutes is required to stabilize within
Support line (781) 890-7440
8
Model IRC350
Version 4.0
06/08