MICS-5131 ETC2 [List of Unclassifed Manufacturers], MICS-5131 Datasheet - Page 3

no-image

MICS-5131

Manufacturer Part Number
MICS-5131
Description
Ethanol Gas Sensor
Manufacturer
ETC2 [List of Unclassifed Manufacturers]
Datasheet
Cross Sensitivity:
The following figure illustrates the MiCS-5131 cross
sensitivity to CH
Response time:
A typical response curve under laboratory conditions is
shown in figure 5. The test gas is 60ppm of C2H5OH and
the sensing resistance reaches 90% of its final change in
about one second.
MiCS-5131 Product Data Sheet
Fig. 4: Sensitivity to CO, H2, CH4, and C2H5OH at
25°C and 50% RH.
0.1
1
10
Fig. 5: Response time to 60ppm C2H5OH
1.2
0.8
0.6
0.4
0.2
1
0
-10
CH4
CO
C2H5OH
H2
4
, CO, C
-5
concentration [ppm]
2
H
5
100
time [second]
OH and H
0
2
.
5
1000
10
Page 3 of 5
P
M i c r o C h e m i c a l S y s t e m s
Temperature and Humidity
dependence:
Temperature and humidity also affect the resistance value
of the sensor. Humidity is water (H2O) in gas phase, which
reacts with the sensing layer like a reducing gas.
Increasing humidity causes a decrease of the sensing
resistance. As for the temperature, the effect is the same
as for the humidity, i.e. decreasing resistance with
increasing
coefficient is due to the semiconductor properties of the
sensing layer material.
R O D U C T
Fig. 6: Temperature dependence of baseline
resistance Rair for 20%, 50%, 70% and 90% RH.
0.1
10
1
temperature.
0
D
20
A T A
This
Temperature [° C]
40
negative
Doc NO: 0178 Rev. B
S
H E E T
60
temperature
Rs/Ro 20%
Rs/Ro 50%
Rs/Ro 70%
Rs/Ro 90%
80

Related parts for MICS-5131