AD7569 Analog Devices, AD7569 Datasheet - Page 19

no-image

AD7569

Manufacturer Part Number
AD7569
Description
LC2MOS Complete/ 8-Bit Analog I/0 Systems
Manufacturer
Analog Devices
Datasheet

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
AD7569JN
Manufacturer:
ADI/亚德诺
Quantity:
20 000
Part Number:
AD7569JNZ
Manufacturer:
SHARP
Quantity:
6 228
Part Number:
AD7569JNZ
Manufacturer:
ADI/亚德诺
Quantity:
20 000
Part Number:
AD7569JP
Manufacturer:
AD
Quantity:
5 510
Part Number:
AD7569JP
Manufacturer:
ADI
Quantity:
244
Part Number:
AD7569JP
Manufacturer:
Analog Devices Inc
Quantity:
10 000
Part Number:
AD7569JP
Manufacturer:
ADI/亚德诺
Quantity:
20 000
Part Number:
AD7569JPZ
Manufacturer:
Analog Devices Inc
Quantity:
10 000
Part Number:
AD7569JPZ
Manufacturer:
ADI/亚德诺
Quantity:
20 000
Part Number:
AD7569JPZ-REEL
Manufacturer:
Analog Devices Inc
Quantity:
10 000
Part Number:
AD7569JRZ
Manufacturer:
Renesas
Quantity:
111
Part Number:
AD7569JRZ
Manufacturer:
ADI/亚德诺
Quantity:
20 000
Part Number:
AD7569KNZ
Manufacturer:
TI
Quantity:
5 925
TARE FUNCTION FOR WEIGH SCALE—AD7569
The infinite sample-and-hold just outlined can also form the ba-
sis of a circuit to provide a tare function for a weigh scale sys-
tem. Figure 30 shows a circuit for a weigh scale system. It
incorporates a tare function using a simple circuit based on the
AD7569.
The AD587, along with the 2N6285, provides a buffered +10 V
reference to supply the low impedance load cell transducer. The
load cell output is amplified by the AD624 precision instrumen-
tation amplifier with gain adjustment provided by R1. The out-
put of the AD624 is applied to the noninverting input of a unity
gain differential summing amplifier that uses the AD707, a high
precision op amp with low drift. The AD707 feeds a 3 1/2 digit
REV. B
Figure 30. Weigh Scale System with Tare Function
–19–
panel meter module that converts the signal for digital readout.
The input signal to the panel meter is also applied to the analog
input of the AD7569 for the tare function. When the tare switch
(S1) is closed, a tare cycle commences and V
held infinitely at V
inverting input of the differential amplifier and forces its output
to zero. Thus, the tare function is used to give a readout of zero
for any undesired weight, such as a box, when only the item
placed in it is to be weighed. The tare function can also be used
in calibrating the system, to cancel out offset errors due to the
load cell, AD624 and differential amplifier.
The AD7569 offers many advantages in the system outlined,
such as: simple, low cost circuit—no need for microprocessor,
software, etc.—and low power consumption.