KMD15-1212 TDK Corporation, KMD15-1212 Datasheet - Page 13

PWR SPLY MEDICAL +12V/-12V 15W

KMD15-1212

Manufacturer Part Number
KMD15-1212
Description
PWR SPLY MEDICAL +12V/-12V 15W
Manufacturer
TDK Corporation
Series
KMr
Type
Medicalr
Datasheets

Specifications of KMD15-1212

Power Supply Type
Switching (Closed Frame)
Voltage - Output
±12V
Number Of Outputs
2
Power (watts)
15W
Applications
Medical
Voltage - Input
90 ~ 264VAC
Mounting Type
Through Hole
1st Output
12 VDC @ 625mA
2nd Output
-12 VDC @ 625mA
Size / Dimension
2.52" L x 1.8" W x 0.93" H (64mm x 45.6mm x 23.5mm)
Power (watts) - Rated
15W
Operating Temperature
-25°C ~ 70°C
Efficiency
80%
Approvals
CE, EN, IEC, UL
Line Regulation
0.5%
Load Regulation
1%
Power Supply Output Type
Fixed
No. Of Outputs
2
Output Voltage
12V
Output Current
625mA
Power Rating
15W
Input Voltage
90V AC To 264V AC / 100V DC To 375V DC
Brand/series
KM Series
Configuration
Encapsulated
Operation
Switching
Output
-12 VDC @ 62.5 mA to 0.625 A, +12 VDC @ 62.5 mA to 0.625 A
Power, Output
15 W
Primary Type
AC-DC
Special Features
High Efficiency
Voltage, Input
90 to 264/100 to 375 VAC/VDC
Voltage, Output
-12, +12 VDC
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Lead Free Status / RoHS Status
Lead free / RoHS Compliant, Lead free / RoHS Compliant
Other names
285-1759
KMD151212

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
KMD15-1212
Manufacturer:
TDK-LAMBDA AMERICAS INC
Quantity:
135
Part Number:
KMD15-1212
Manufacturer:
TDK-LAMBDA
Quantity:
253
Special Instructions for medical applications (IEC/EN/UL/CSA 60601-1)
These products are designed for continuous operation within an overall enclosure, and must be mounted
such that access to the mains terminals is restricted. See clause 16, IEC60601-1, EN60601-1 and UL/CSA
60601-1.
Input Connection - KMx15 –xx
These products are NOT suitable for use in the presence of flammable anaesthetic mixtures with air or with
oxygen, or with nitrous oxide.
These products are classed as ordinary equipment, type B and have a reinforced barrier between input and
output to IEC60601-1, EN60601-1 and UL/CSA 60601-1. They are NOT protected against the ingress of
water.
Reference should be made to local regulations concerning the disposal of these products at the end of their
useful life.
Environmental Specifications:
1
2
3
4
5
6
7
8
9
Overvoltage Category
Product HANDBOOK
Pollution Degree
Pin
Material Group
Temperature
Description
Orientation
Pressure
Humidity
Vibration
Altitude
Shock
Class
Use
No pin
AC (L)
AC (N)
-DC out
No pin
No pin
+DC out
No pin
No pin
Single O/P
-25° C - +70° C (derating at 2.5%/° C above 50° C)
No Pin
AC (L)
AC (N)
-DC out
GND
No pin
+DC out
No pin
No pin
Dual O/P
20 - 95% RH, non-condensing
23.52m/s
KM15 Series
No Pin
AC (L)
AC (N)
GND for +5V
+5V
-DC out
No pin
GND
+DC out
70kPa - 106kPa
-200m - 3000m
Triple O/P
Operation
2
Indoor
(10-55Hz): constant sweep 1min X, Y, Z for 1 hour
<196.1m/s
IIIb
2
II
II
2
(20G)
20 - 95% RH, non-
54kPa - 106kPa
-40° C - +100°C
-200m - 5000m
condensing
Storage
ALL
-