NFS40-7924 Emerson Network Power, NFS40-7924 Datasheet - Page 2

no-image

NFS40-7924

Manufacturer Part Number
NFS40-7924
Description
PS SWITCHER MEDICAL 24V 40W
Manufacturer
Emerson Network Power
Series
NFS40r
Datasheet

Specifications of NFS40-7924

Rohs Status
RoHS non-compliant
Voltage - Output
24V
Number Of Outputs
1
Power (watts)
40W
Applications
Medical
Power Supply Type
Switching (Open Frame)
Voltage - Input
85 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
24 VDC @ 1.6A
Size / Dimension
5" L x 3" W x 1.2" H (127mm x 76.2mm x 30.5mm)
Power (watts) - Rated
40W
Operating Temperature
0°C ~ 70°C
Efficiency
70%
Approvals
CSA, UL, VDE
Line Regulation
±0.2%, ±1%
Load Regulation
±2%
3rd Output
-
2nd Output
-
4th Output
-

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
NFS40-7924
Manufacturer:
ARTESYN
Quantity:
1 000
Electrical Specifications Continued
Environmental Specifications
All specifications are typical at nominal input, full load at 25
EMC Charateristics
General Specifications
Conducted emissions:
Radiated emissions:
ESD air:
ESD contact:
Surge:
Fast transients:
Radiated immunity:
Conducted immunity:
Thermal performance:
Relative humidity:
Altitude:
Vibration (See Note 11):
Hold-up time:
Efficiency:
Isolation voltage:
Switching frequency:
Approvals and standards:
(See Note 12)
Weight:
MTBF demonstrated:
EN55022, FCC part 15
EN55022, FCC part 15
EN61000-4-2, level 3
EN61000-4-2, level 4
EN61000-4-5, level 3
EN61000-4-4, level 3
EN61000-4-3, level 3
EN61000-4-6, level 3
110 Vac
230 Vac
110 Vac, 230 Vac
Input/output
Input/chassis
VDE0750, IEC601, EN60601-1, UL2601, CSA C22.2 No. 125
MIL-HDBK-217E
Operating, see curve
Non-operating
0 - 50 °C ambient temperature,
convection cooled
0 - 50 °C ambient temperature
+50 °C to +70 °C ambient
Peak (30 seconds)
Non-condensing
Operating
Non-operating
5 - 500 Hz
°
C unless otherwise stated
Level A
Level A
Perf. criteria 1
Perf. criteria 1
Perf. criteria 1
Perf. criteria 1
Perf. criteria 2
Perf. criteria 2
18 ms
132 ms
70% typical
4000 Vac
1500 Vac
20 - 110 kHz
270 g (9.6 oz)
170,000 hours
0° C to +70 °C
-40 °C to +85 °C
40 W
50 W @ 20 CFM
Derate linearly to 50% load
60 W
5 to 95% RH
10,000 feet max.
40,000 feet max.
0.75 G peak
Embedded Power for
Business-Critical Continuity
Rev.06.9.09_118