I1010 Fluke Electronics, I1010 Datasheet - Page 7

no-image

I1010

Manufacturer Part Number
I1010
Description
CURRENT CLAMP AC/DC 1000A
Manufacturer
Fluke Electronics
Type
AC/DC Clamp, Hall Effectr
Datasheets

Specifications of I1010

Mfg Application Notes
Beat the Book, Troubleshooting, Appl Notes
Style
Pincer
Ac Current Range
1 ~ 600A
Bandwidth
10kHz
Cable Length
62.992" (1600.00mm)
Dc Current Range
1 ~ 1000A
Max Conductor Size
1.181" (30.00mm)
Termination
Banana Plug, Double, Sheathed
Ratings
CAT III 600V
For Use With
FLUKE-116 - MULTIMETER HVAC TRUE RMS614-1063 - COMBO KIT DMM & ACCESSORY614-1062 - DMM TRUE RMS 50000CNT W/BCKLGHT614-1016 - INSULATION MULTIMETER .01M-2G614-1015 - INSULATION MULTIMETER 0.1M-600M
Lead Free Status / RoHS Status
Lead free by exemption / RoHS compliant by exemption
Other names
617735
Operating Temperature
Storage Temperature
Relative Humidity
Altitude
Battery Type
Battery Life
Safety
Current x Hz Product
Output Signal
Working Voltage
Maximum Conductor Size
Load Impedance
Temperature Coefficient
Size
Weight
Specified Current
Range:
Usable Current
Range:
DC Accuracy (zero
adjusted, conductor
centered)
AC Accuracy
Bandwidth
*
Specifications
Operating
Non-operating
With a true-rms voltmeter, the minimum ac current is limited to the low
end of the specified mV ac range.
1 A – 400 A ac rms *
1 A – 400 A dc
0.5 A – 400 A
3.5 % + 0.5 A (0 A –
400 A)
3.5 % + 0.5 A, 45 Hz –
400 Hz, Crest Factor
(0 A – 400 A)
3 kHz
3.
-10 °C to +50 °C (14 °F to 122 °F)
-20 °C to +60 °C (-4 °F to +140 °F)
0 % to 95 % (0 C to 30 C)
0 % to 75 % (30 C to 40 °C)
0 % to 45 % (40 C to 50 °C)
0 m – 2000 m
0 m – 12000 m
60 hrs typical (continuous with alkaline)
s
1 mV per amp dc or ac
600 V rms, CAT. III maximum at input
1 ea. 30 mm (1.18 in.) diameter
2 ea. 25 mm (0.98 in.) diameter
+/-(0.05 x accuracy per °C (0 °C – 18 °C,
28 °C – 50 °C)
209 mm x 78 mm x 48 mm
(3.09 in. x 8.21 in. x 1.87 in.)
0.5 kg (17.6 oz.)
9 V (alkaline) NEDA 1604 IEC 6F22
1 M , 100 pF
240,000
Double Insulation, 600 V rms, CAT. III
i410
1 A – 600 A ac rms *
1 A – 1000 A dc
0.5 A – 1000 A
2.0 % + 0.5 A (0 A –
1000 A)
2.0 % + 0.5 A, 45 Hz
– 400 Hz, Crest
Factor
3.0 % + 0.5 A, 400 Hz
- 2 kHz sine wave.
(0 A – 600 A)
10 kHz
i1010
3.